sc100 Operation
4.5.2 Sample Calibration
Step
Select
Menu Level/Instructions
Confirm
1
—
Remove the probe from the process and clean it
—
2
—
Obtain a sample solution with a known value.
—
3
MAIN MENU
—
4
SENSOR Setup
5
Highlight the appropriate sensor if more than one sensor is attached.
6
CALIBRATE
7
SAMPLE CAL
8
Choose from the available Output Mode options (Active, Hold, or Transfer). An output
that has been placed in Hold or Transfer status will be automatically released when the
calibration is complete.
9
—
Move the clean probe to the sample.
10
The display will prompt for entry of a salinity value. Use the
UP
and
DOWN
keys to
change the digit that is flashing. Press the
LEFT
key to move to another digit.
11
—
The display will show “Press Enter When Stable” and the current DO and temperature
readings. When
ENTER
is pressed or when the reading has been accepted as stable,
the display will change to an entry screen. If
ENTER
is not pressed, the sensor will
determine when the reading is stable. If it is not stable within 45 minutes, the display
will show “Unable to Calibrate”.
12
When a stable reading has been accepted, the display will show “Sample Value” and
an area for entry of the sample value. Edit the displayed reading.
13
—
The display will show a response from
.
14
—
Return the probe to the process.
—
15
The display will show “Press Enter when Stabilized” and the current DO and
temperature readings. When
ENTER
is pressed or when the reading has been
accepted as stable, the display will change to an entry screen. If
ENTER
is not pressed,
the sensor will determine when the reading is stable. Two to three minutes for the
reading to stabalize is typical, however, if it has not stabilized after 45 minutes, the
display will show
“Unable to Calibrate”.
16
—
When a stable reading has been accepted, the display will show “Sample Cal” and an
area for entry of the value obtained from the Winkler Titration.
—
17
Main Menu or Main Measurement Screen
—
客服:400 688 9809
Содержание 5740sc
Страница 1: ...Visit http www hach com 客服 400 688 9809 ...
Страница 6: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 12: ...12 Installation Figure 4 Sensor Installation Examples 客服 400 688 9809 ...
Страница 22: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 34: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 38: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 40: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 44: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 48: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 50: ...Visit us at www hach com 客服 400 688 9809 ...