21
sc100 Operation
Table 4 Calibration Response
Calibration Response
Explanation
Complete
Indicates the calibration is complete and stored in the instrument.
Cal Fail, Offset High
Indicates the air calibration has failed due to an excessively high calculated gain
value. Repeat the calibration.
Cal Fail, Offset Low
Indicates the air calibration has failed due to too low of a calculated gain value.
Repeat the calibration.
Cal Fail, Unstable
Indicates the air calibration has failed because the readings did not stabilize during the
maximum allowed calibration time interval. Repeat the calibration.
4.5.3 Concurrent Calibration of Two Sensors
Step
Select
Menu Level/Instructions
Confirm
1
—
Begin a calibration on the first sensor and proceed to “Wait to Stabilize” is displayed.
2
—
Select LEAVE.
3
—
The display will return to the main measurement screen. The reading for the sensor
currently being calibrated will be flashing.
—
4
—
Begin the calibration for the remaining sensor and proceed to “Wait to Stabalize” is
displayed.
5
—
Select LEAVE.
6
—
The display will return to the main measurement screen and the reading for both
sensors will be flashing.
7
—
To return to the calibration of either sensor, press
MENU
, select Probe Setup and
ENTER
. Select the appropriate sensor and
ENTER
.
客服:400 688 9809
Содержание 5740sc
Страница 1: ...Visit http www hach com 客服 400 688 9809 ...
Страница 6: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 12: ...12 Installation Figure 4 Sensor Installation Examples 客服 400 688 9809 ...
Страница 22: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 34: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 38: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 40: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 44: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 48: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 50: ...Visit us at www hach com 客服 400 688 9809 ...