10
Installation
Figure 2
Hard-wiring the Sensor
3.2 Connecting the Sensor to the sc1000
1.
Unscrew the protective cover on the controller socket. Retain the protective cover for
the sensor. When the sensor is removed, re-fit the protective cover.
2.
Push the connector into the socket. During this process pay attention to the lugs on
the connector.
3.
Hand-tighten the union nut.
Note: Do not use the middle connection for the sensors, which is reserved for the display module.
Table 1 Wiring the Sensor at Terminal Block J5
Terminal Number
Terminal Designation
Wire Color
J5-1
Data (+)
Blue
J5-2
Data (–)
White
J5-3
Service Request
No Connection
J5-4
+12 V dc
Brown
J5-5
Circuit Common
Black
J5-6
Shield
Shield (grey wire in existing quick-disconnect fitting)
1
1
+
DATA
+
DATA
+
OUT 2
+
OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+
V
+
V
+
OUT 1
+
OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY 3
RELAY 2
RELAY 1
1
1
+
DATA
+
DATA
+
OUT 2
+
OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+
V
+
V
+
OUT 1
+
OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
U5
U9
S1
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
1
1
+
DATA
+
DATA
+
OUT 2
+
OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+
V
+
V
+
OUT 1
+
OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY 3
RELAY 2
RELAY 1
1
1
+
DATA
+
DATA
+
OUT 2
+
OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+
V
+
V
+
OUT 1
+
OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
U5
U9
S1
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
J6
J5
J5
J5
J5
Disconnect
Power
From Probe
客服:400 688 9809
Содержание 5740sc
Страница 1: ...Visit http www hach com 客服 400 688 9809 ...
Страница 6: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 12: ...12 Installation Figure 4 Sensor Installation Examples 客服 400 688 9809 ...
Страница 22: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 34: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 38: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 40: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 44: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 48: ...Visit us at www hach com 客服 400 688 9809 ...
Страница 50: ...Visit us at www hach com 客服 400 688 9809 ...