Option
Description
SAMPLE HOLDER FLUSH
Sets the relay to on when the sample holder flushes.
DECOMISSIONING
Sets the relay to on when a decommissioning cycle is started.
MARK END OF MEASURE
Sets the relay to on at the end of a measurement.
Set the error hold mode
1.
Push
menu
and select SETUP SYSTEM>SETUP OUTPUTS>ERROR HOLD MODE.
2.
Select an option.
Option
Description
HOLD OUTPUTS
Holds the outputs at the last known value when communications are lost.
TRANSFER OUTPUTS
Switches to the transfer mode when communications are lost. The outputs transfer
to a pre-defined value.
View data
1.
Select VIEW DATA.
2.
Select an option.
Option
Description
ANALYZER DATA
Shows analyzer status information (refer to
).
MEASUREMENT DATA
Shows measurement information (refer to
LOG DATA
Selects the data log and the event log. DATA LOG—shows the measurement
values. Select start time, number of hours and number of readings. EVENT LOG—
shows all analyzer information. Select start time, number of hours and number of
readings.
Table 2 Analyzer data
Element
Definition
CELL 1 TEMP
CELL 2 TEMP
Temperature of the colorimeter cell block heater (ideally 40 to 45 °C (104 to 113 °F))
REAGENT TEMP
Reagent temperature before it enters the colorimeter
AMBIENT TEMP
Air temperature inside the electronics area
AIR PRESS
Reagent air pressure in the reagent bottles (ideally 3.95 to 4.05 psi)
LED DUTY CYCLE
2
Depends on condition of colorimeter cell and age of analyzer (typically 7,200 to
40,000 counts)
HEATER DUTY CYCLE
1
Percentage of time the colorimeter heater is on to maintain a constant 40 °C (104 °F)
SAMPLE FLOW
Approximate sample flow into the sample holder during the sample holder flush cycle.
SAMPLE PRESS 1
Sample pressure before the internal filter for channel 1 (ideally 2 to 4.5 psi depending
on incoming sample pressure)
SAMPLE PRESS 2
Sample pressure before the internal filter for channel 2 (recommended at 2 to 4.5 psi
in relation to the incoming sample pressure)
REAGENT 1
Level of reagent remaining
2
Information is available for both sample cells
16
English
Содержание 5500sc
Страница 1: ...DOC023 97 80474 5500sc Ammonia Monochloramine 05 2017 Edition 4 Operations Fonctionnement Operaciones...
Страница 2: ...English 3 Fran ais 20 Espa ol 39 58 2...
Страница 58: ...58 60 62 63 73 58...
Страница 59: ...pH SD 1 SD 1 SD 1 3 LED 2 1 4 SD 1 59...
Страница 60: ...1 1 2 2 1 6 Diag DIAG TEST MENU 2 RIGHT LEFT 7 Cal CALIBRATE MENU 3 UP DOWN 8 Back 4 Home 9 Menu 5 Enter 3 4 60...
Страница 74: ......
Страница 75: ......