![Hach 2200 PCX Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/hach/2200-pcx/2200-pcx_basic-user-manual_3152422008.webp)
Mechanical installation
Attach the instrument to the wall
W A R N I N G
Personal injury hazard. Make sure that the wall mounting is able to hold 4 times the weight of the
equipment.
This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Use of this instrument at an
altitude higher than 2000 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.
Attach the instrument upright and level on a flat, vertical surface. Refer to
Figure 3
for dimensions.
Mounting hardware is supplied by the user. Make sure that an electrical outlet is available that is
above flood stage areas. If purchased with the instrument, install the standard water weir flow
controller so that the sensor outlet of the instrument is lower than the water weir overflow. Refer to
Figure 1
on page 6.
Figure 3 Installation dimensions
Plumbing
D A N G E R
Fire hazard. This product is not designed for use with flammable liquids.
8
English
Содержание 2200 PCX
Страница 1: ...DOC023 97 80464 2200 PCX 04 2018 Edition 2 Basic User Manual Manuel d utilisation de base...
Страница 2: ...English 3 Fran ais 25 2...
Страница 49: ......