BAROMETRIC PRESSURE SENSOR

The instrument internal barometric pressure
sensor is calibrated at the factory, and normally
requires no further attention. However, you may
w i s h   t o   c a l i b r a t e   i t   a g a i n s t   y o u r   o w n
instrumentation, or simply check the instrument
for accuracy.
Select 

Troubleshooting

Pressure Calibration

from the Hach Ultra 3658 program, and enter the
current atmospheric pressure, in mbars, in the
calibration pressure entry box.

GAS SENSOR

The sensor can be calibrated in a liquid solution
w i t h   a   k n o w n   c o n c e n t r a t i o n   o f   C O

2

  ( w e

recommend using the ORBISPHERE High
Precision Calibration Kit), or in pure CO

2

 gas at

atmospheric or elevated pressure.
Before calibrating check that the sensor
membrane surface is tight, smooth and wrinkle
free. Select the calibration medium using the
Hach Ultra 3658 program.
Ensure that normal measurement mode is
selected in the Measurement Mode menu of the
Hach Ultra 3658 program
Calibration must only be carried out once the TC
sensor is giving a stable measurement. Expose
the membrane to the calibration sample until this
stable reading is obtained. This usually takes
about 5 minutes. 
For calibration it is important that the sensor's
membrane is at the same temperature as the
calibration sample. Therefore, allow some sample
to pass through the flow chamber for 3 minutes
before calibrating.
If calibrating in gas, ensure that the flow chamber
and sensor components are completely dry.
Remove and blow dry the flow chamber, and dry
the sensor head surface with a clean soft tissue. 
Ensure that the purge cycle is running correctly by
placing the purge gas exit tubing into water. This
rate should be more than four bubbles during the
purge which lasts for 4 seconds. This purge cycle
is repeated every 20 seconds.

Note :

When press the 

CAL

 button, remember that this 

button may have been locked out by the Hach 
Ultra 3658 program to prevent an accidental 
reset.

In all calibration methods, calibration is successful
or in error, as per the following definitions:

If the instrument makes two consecutive 
measurements with less than 1% 
difference, it calibrates against this 
stabilized value. The LCD then displays 
the message 

[CAL]

 to indicate 

calibration has been successfully 
completed

If the calibration does not complete, the 
LCD displays the message 

[Err]

. The 

reason for this calibration error is either 
that the measured gas partial pressure is 
under 5 mbar, or that a wrong instrument 
key was pressed during the calibration 
steps.

Calibration in a Liquid Solution

The sensor can be calibrated in a liquid solution
that has a known concentration of CO

2

.

Calibration

Содержание ORBISPHERE 3658

Страница 1: ...Operator Quick Guide ORBISPHERE 3658 Revision A 28 04 2008 ...

Страница 2: ... performs a series of start up tests before switching to measurement mode MEAS places the instrument in measurement mode CAL calibrates the analyzer against a reference sample This button can be locked out from the Hach Ultra 3658 PC program STO stores a measurement value into memory The Ø keys toggle between gas concentration and temperature measurement displays in measurement mode increases or d...

Страница 3: ...hing the instrument back ON The LCD displays the message Sto for about one second Gas concentration measurements are displayed for about two minutes After two minutes the instrument displays the sample number starting at 000 then the gas concentration measurement value followed by to indicate the measurement is being stored This storage sequence repeats automatically at the rate specified To end a...

Страница 4: ...time stamp and a sample description Altering Sample Point Description For help in identifying the locations of various sampling points that are stored you may choose the Sampling Point Description command from the Logger menu The measurement values to be placed in positions 0 through 499 can be described however you wish Double click on a particular position or click Modify then type a description...

Страница 5: ... oldest samples are lost off the left of the chart When Continuous is not checked the chart stops displaying new results after reaching the 10 division and all subsequent measurements are lost Click the up down pointers for each measurement variable GAS TEMPERATURE and PRESSURE to change the scaling of that value on the chart The display of each measurement variable may be turned on or off by choo...

Страница 6: ...e valid for CO2 gas measurement with this analyzer To ensure the analyzer is correctly configured choose Membrane to bring up the box which reveals the membrane models available Membrane 29561A should be the only membrane available for selection SETTING SAMPLING INTERVALS The analyzer can perform as a standalone data acquisition device automatically recording gas measurements with the date and tim...

Страница 7: ...tion AUTOMATIC SHUTDOWN Choose Automatic Shutdown Status from the Configuration menu to activate the Automatic Shutdown feature If you select Enable the instrument switches off automatically after 10 minutes of inactivity thereby economizing battery power MEASUREMENT MODE Choose Measurement Mode to enable maximum measurement mode for sampling in bottles and cans In Maximum measurement mode the ins...

Страница 8: ... same temperature as the calibration sample Therefore allow some sample to pass through the flow chamber for 3 minutes before calibrating If calibrating in gas ensure that the flow chamber and sensor components are completely dry Remove and blow dry the flow chamber and dry the sensor head surface with a clean soft tissue Ensure that the purge cycle is running correctly by placing the purge gas ex...

Страница 9: ... of pure CO2 gas to the inlet of the flow chamber and adjust the gas flow to one bubble per second when the exit tube is immersed in water Press the CAL button A brief clearing message appears Press CAL again within a 3 second period Calibration is successful or in error as defined earlier Calibration in CO2 at Elevated Pressure This method requires an accurate pressure gauge connected to the exit...

Страница 10: ... should darken DISPLAY TEST Choosing the Display Test option lets you perform a one way communication between computer and instrument Type a number in the Number box you may also select one of three units positions for the LCD s rightmost indicator bar as well Then choose Send The number and indicator bar placement should appear on your instrument LCD CLOCK SETTINGS Choose the Clock settings optio...

Страница 11: ... Membrane not clean Clean system Incorrect solubility curve Change solubility Incorrect calibration Recalibrate High readings Membrane is not at sample temperature Allow sample to flow past the membrane to make sure the membrane is at sample temperature before mak ing measurements Incorrect solubility curve Change solubility Incorrect calibration Recalibrate Message Meaning Pur This message appear...

Страница 12: ...nsor Maintenance It is recommended to leave the sensor in place in the instrument when changing the membrane To remove the membrane follow the steps below To avoid any liquid coming into contact with the TC sensor chip via the purge path check that the flow chamber is completely empty of any liquid sample Then remove the flow chamber from the sensor by turning the flow chamber locking nut counter ...

Страница 13: ...downwards Take a few membranes out of the storage box Using tweezers included in the kit pick up one membrane from the stack and gently place it on the sensor tip Make sure it is centered If a sensor mask is used place it directly on top of the membrane Note Distinguish the membrane from the protection paper The membrane is transparent translucent The protection paper is opaque The membrane diamet...

Страница 14: ...the protection cap with a new one Tighten the protection cap finger tight Then complete the process using the tool provided in the maintenance kit Insert into each of the four holes in turn and tighten as far as possible Tighten each hole only once Note The grille inside the protection cap should be free to move during tightening Therefore and to avoid damage to the membrane do not touch the grill...