3624 ProBrix Plus - Installation
Operator Manual
ORBISPHERE
1.6 Sensor Installation
The Brix/Diet, O
2
, and CO
2
sensors are mounted and connected inside the sampling
module (see
). For detailed information on installation and
preparation, please refer to the appropriate accompanying
Sensor Maintenance and
Installation Manuals
.
Once installed, the CO
2
sensor requires additional attention.
1.6.1 CO
2
Sensor Purge Gas Connection
Before installing the CO
2
sensor, make sure that:
1) The instrument is switched ON
2) The purge gas is connected (see following instructions) and supplied at 2-bar
gauge pressure
CAUTION:
Do not place the CO
2
sensor into a liquid sample until the above is true, as without a
constant supply of dry purge gas, water could condense inside the measuring chamber
and cause damage to the thermal conductor chip.
To ensure that the purge gas is dry and maintained at the correct pressure, a model
29089 regulator and filter (see
) is mounted inside the sampling module.
Fig 1-14:
29089 Gas Pressure Regulator and Filter
Содержание ORBISPHERE 3624
Страница 1: ...Operator Manual ORBISPHERE 3624 Revision F 03 10 2008 ...
Страница 2: ......
Страница 14: ...8 of 96 Manual Overview 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 32: ...26 of 96 Installation 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 60: ...54 of 96 PC Program Setup 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 68: ...62 of 96 Calibrations 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 82: ...76 of 96 Accessories and Attachments 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 98: ...92 of 96 Glossary 3624 ProBrix Plus Operator Manual ORBISPHERE ...
Страница 101: ...3624 ProBrix Plus User Notes 95 of 96 Operator Manual ORBISPHERE User Notes ...
Страница 102: ......