24
Operation
4.10.3 Clear the buffer
To clear the buffer:
1.
Press the
MODE
key three times on the Main screen. The bottom line displays the
number of records in the Buff Count field (refer to
2.
Press and hold the
MODE
key for one second, until two beeps are heard.The Buffer
count will be displayed on the Buffer Options screen.
3.
Press the
START/STOP
key to delete the buffer records. The Delete Buffer screen is
displayed (refer to
4.
Press the
START/STOP
key to confirm the deletion. The Main screen displays with the
buffer count shown at the bottom line.
Note:
The buffer count should be equal to zero.
Figure 14 Buffer options screen
Figure 15 Delete Buffer screen