Maintenance menu
Select
Maintenance
from the main menu.
Syringe activation
Do a check of the syringe. Make sure the syringe fills and empties correctly.
1.
From the maintenance menu, push
Syringe activation
.
Option
Description
Fill
Fills the syringe with titrant solution. The process stops automatically when the syringe is
full.
Empty to bottle
Discards the contents of the syringe into the titrant bottle.
Empty to beaker
Discards the contents of the syringe into the beaker. Make sure that the tube from the
outlet port of the syringe is inside the beaker.
Stop
Stops the operation.
Syringe replacement
To replace the syringe, select
Syringe replacement
from the maintenance menu. Obey the
instructions on the display. Refer to
on page 8.
Pump activation
This option only applies to instruments with pumps installed. Do a check of the pump. Make sure the
pump fills and empties correctly.
1.
From the maintenance menu, push
Pump activation
.
Option Description
Start
Starts the pump. The reagent is pumped through the tubes into the beaker. Make sure that the tube
from the outlet port of the pump is in the beaker.
Stop
Stops the operation.
2.
Push
Toggle
to change to the second pump.
Pump cassette replacement
D A N G E R
Electrocution hazard. Remove power from the instrument before this procedure is started.
This option only applies to instruments with pumps installed.
To replace the pump cassette, refer to the illustrated steps that follow and obey the instructions on
the display.
20
English
Содержание TitraLab KF1000 Series
Страница 21: ...English 21 ...
Страница 130: ...130 Português ...
Страница 151: ...Čeština 151 ...
Страница 172: ...172 Dansk ...
Страница 237: ...Svenska 237 ...
Страница 322: ...322 Română ...
Страница 343: ...Türkçe 343 ...
Страница 364: ...364 Slovenský jazyk ...
Страница 385: ...Slovenski 385 ...
Страница 406: ...406 Hrvatski ...
Страница 428: ...428 Ελληνικά ...
Страница 431: ......