background image

Figure 1  Keypad description

1   

Enter: Saves the setting and exits the current
screen to the CONFIGURE menu

4   

Delete: Removes a sensor from the selection

2   

Cancel: Exits the current screen to the
CONFIGURE menu without saving the setting

5   

UP and DOWN arrows: Moves the sensors up or
down the list

3   

Add: Adds a new sensor to the selection

Startup

Add a sensor

Note: PROGNOSYS is not available for all sensors. Only newer sensor models can use the PROGNOSYS
function.
Note: Make sure that an RTC or a PROGNOSYS communication card is installed in the sc1000 sensor module.

When an RTC or PROGNOSYS communication card is installed, PROGNOSYS files are available
for different sc sensors. Do the steps that follow to add a sensor. Refer to 

Figure 2

.

1.

Connect the controller. Refer to the controller documentation.

2.

Select an option.

Option

Description

For
RTC/PROGNOSYS
card

Select MAIN MENU>RTC MODULES/PROGNOSYS>RTC
MODULES>RTC>CONFIGURE>SELECT SENSOR

For PROGNOSYS
card

Select MAIN MENU>RTC
MODULES/PROGNOSYS>PROGNOSYS>CONFIGURATION>PROGNOSYS>SELECT
SENSOR

3.

Push 

Add

. A list with all network connections opens.

4.

Select the applicable sensor for the RTC or PROGNOSYS module and push 

Enter

. The sensor is

shown in the sensor list.

Note: Sensor names with black font are available for an RTC module. Sensor names with red font are not
available for an RTC module. A sensor name identified with a "(p)" is available for PROGNOSYS.

5.

Push 

Add

 to add more sensors from the list.

Sensors that are selected before are shown in gray. Refer to 

Figure 3

 on page 7 or 

Figure 4

on page 7 to put in order or delete a sensor.

6.

Push 

Enter

 to accept the list.

English

   

5

Содержание PROGNOSYS

Страница 1: ...DOC023 52 90351 PROGNOSYS 05 2014 Edition 7 User Manual ...

Страница 2: ......

Страница 3: ...ensor 5 Sort the sensors RTC modules only 6 Delete a sensor from the list 7 Operation 7 PROGNOSYS indicator bars 7 Display description 8 Measurement value indicator 9 Service indicator 9 See the message lists 9 Configure the general settings 10 Configure the sensor 11 Service and measurement value indicator messages 11 AMTAX sc and PHOSPHAX sc 11 NITRATAX plus sc 13 ANISE sc AISE sc NISE sc 13 SOL...

Страница 4: ...Table of Contents 2 ...

Страница 5: ...y this equipment is not impaired Do not use or install this equipment in any manner other than that specified in this manual Use of hazard information D A N G E R Indicates a potentially or imminently hazardous situation which if not avoided will result in death or serious injury W A R N I N G Indicates a potentially or imminently hazardous situation which if not avoided could result in death or s...

Страница 6: ...acement part Product components Make sure that all components have been received If any items are missing or damaged contact the manufacturer or a sales representative immediately PROGNOSYS can only be used with an sc1000 controller that has an RTC communication card or a PROGNOSYS card installed The PROGNOSYS files for the different sc sensors are kept on the sc1000 controller To use PROGNOSYS wi...

Страница 7: ...or Refer to Figure 2 1 Connect the controller Refer to the controller documentation 2 Select an option Option Description For RTC PROGNOSYS card Select MAIN MENU RTC MODULES PROGNOSYS RTC MODULES RTC CONFIGURE SELECT SENSOR For PROGNOSYS card Select MAIN MENU RTC MODULES PROGNOSYS PROGNOSYS CONFIGURATION PROGNOSYS SELECT SENSOR 3 Push Add A list with all network connections opens 4 Select the appl...

Страница 8: ...ensor 3 Sensor list Sort the sensors RTC modules only The sensor sequence is programmed in the RTC module for the measurement values To sort the sensors in the order specified for the RTC module move the selected sensor with the UP and DOWN arrows Refer to Figure 3 6 English ...

Страница 9: ...o back without changes Operation PROGNOSYS indicator bars If PROGNOSYS is installed two additional horizontal bars show in the top right hand corner of the measurement screen The top bar is the measurement value indicator The bottom bar is the service indicator If no bars show up refer to Troubleshooting on page 17 The position of the cursor and the value in the indicator identifies the sensor sta...

Страница 10: ...tor is not a linear degrading indicator because of the sensor special model based calculation Measurement zone Range Green zone 100 to 75 Yellow zone 75 to 50 Red zone 50 to 0 The exact status shows with an additional vertical cursor A maximum of four measurement values show if multiple sensors are installed on the sc1000 The indicators are shown for each installed sensor in the top right hand cor...

Страница 11: ...d the color automatically changes from red to green Service indicator The service indicator shows the maintenance status of the sensor The bottom bar shows the number of days that remain until a maintenance task must be completed The date and time for the next maintenance task e g cleaning or calibration is calculated Service tasks with no countdown show immediately and change the color from green...

Страница 12: ...ervice indicator Time in days clean Solu days 1 day Reagent days 6 days Configure the general settings Configure service messages to send emails to the manufacturer telemetry service and to a maximum of four freely configured email address es This email gives information about important changes in the measurement indicator and about pending maintenance tasks In addition the length of the YELLOW PH...

Страница 13: ...tor bar flashes when the value decreases below the given value Range 0 to 200 days default 0 days Service and measurement value indicator messages W A R N I N G For safe use of the instrument obey the precautions and the instructions in the sensor documentation This list gives a general overview for the service and measurement value indicator messages Refer to the sensor documentation for a more d...

Страница 14: ...ation solution on time Set the counter when the maintenance task is completed Electrolyte days AMTAX sc only Shows the days that remain to change the electrolyte in the electrode and the membrane cap Change the electrolyte in the electrode and the membrane cap on time Set the counter when the maintenance task is completed Compressor replacement The time for the compressor is expired For instrument...

Страница 15: ...ntact technical support Wiper position unknown Moist Moisture in the sensor is over limit the desiccant is expired There is possibly a problem with the gasket set for the sensor sleeve or the wiper axle Contact technical support immediately to replace the gasket so that the instrument operates correctly R too high The automatic zeroing is not correct Contact technical support for a zero point cali...

Страница 16: ...calibration is started Replace cartridge The time one year for the cartridge is expired Examine the measurement values and replace the cartridge as soon as possible no contact Ref1 Problem with the contact between sensor and reference system Remove the cartridge Examine and clean the contacts Examine the spring action of the contacts in the sensor and replace the spring insert if necessary Contact...

Страница 17: ...n 7 5 C 45 5 F for NH4 Make sure to put the sensor fully in the medium Complete new MX1 or MX2 VC1 or VC2 near to the temperature of the medium during operation NO3 MX2 high Temp diff P1 P2 The temperature difference between the two correction points of a matrix correction MX2 or value correction VC2 for NO3 is more than 7 5 C 45 5 F Select the correction points for MX2 VC2 within a temperature di...

Страница 18: ...ately to replace the gasket so that the instrument operates correctly LED faulty LED intensity is too low Contact technical support Replace gasket The time for the wiper axle gasket cycles is expired Contact technical support to replace for a wiper axle gasket LDO Probe Model 2 Message Possible cause Solution Red amplitude error No sensor cap is installed Refer to the LDO manual to install the sen...

Страница 19: ...d is installed correctly Select MAIN MENU RTC MODULES PROGNOSYS PROGNOSYS ALLOCATION MAP Refer to the communication card installation documentation for more information The bars do not show RED YELLOW or GREEN Only the background color is shown No sensor data is available The RTC PROGNOSYS communication card is not installed correctly Make sure that the RTC PROGNOSYS communication card is installe...

Страница 20: ...fault If the customer fails to notify the supplier the product is considered approved despite the defect Further liability for indirect or direct damages is not accepted If device specific maintenance or inspection work prescribed by the supplier is to be performed within the guarantee period by the customer maintenance or by the supplier inspection and these requirements are not met claims for da...

Страница 21: ...icator7 MEAS INDICAT 7 40183 Unsigned Integer 1 PrognosysServiceStat7 SERVICE STAT 7 40184 Unsigned Integer 1 PrognosysMeasIndicator8 MEAS INDICAT 8 40185 Unsigned Integer 1 PrognosysServiceStat8 SERVICE STAT 8 40186 Unsigned Integer 1 PrognosysMeasIndicator9 MEAS INDICAT 9 40187 Unsigned Integer 1 PrognosysServiceStat9 SERVICE STAT 9 40188 Unsigned Integer 1 PrognosysMeasIndicator10 MEAS INDICAT ...

Страница 22: ...20 English ...

Страница 23: ... color code 8 Bar indicator Bar color code 7 Blink mode 11 C Communication card 4 M Measurement value indicator 9 Message list 9 P PROGNOSYS communication card 4 R RTC communication card 4 S Service indicator 9 21 ...

Страница 24: ......

Страница 25: ......

Страница 26: ...h com www hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2011 2014 All rights reserved Printed in Germany ...

Отзывы: