36
Operation
5.3 [+DEVICEDATA] menu
Instrument parameters for adaptation to the related conditions.
Note:
While the instrument is in this menu, sample pumping continues. If a key is not
pressed for longer than ten minutes, the instrument automatically returns to the display of
the data on operation.
*
If the power supply is interrupted (power failure) both relay contacts are closed.
MENU
Command
Setting
Description
+DEVICEDATA
Δ
CONTRAST
-90 to +90
Reading angle for the display
LANGUAGE
D, GB, NL, F, I, E, PL, S
HEATER
ON, OFF, TIMER (time control)
Default setting: TIMER
Operation of the tube heaters
HEAT.START
(only for [HEATER]:TIMER)
JAN to DEC,
Default setting: OCT
Activation of the heater at the start of the month selected
HEAT.STOP
(only for [HEATER]:TIMER)
JAN to DEC,
Default setting: MRZ (March)
Deactivation of the heater at the end of the month selected
WARNING<
200...800 ml/h in 10 ml steps
Warning relay contact
*
set when the sample volume drops below
the amount set
FAULT<
100...600 ml/h in 10 ml steps
Fault relay contact
*
set when the sample volume drops below the
amount set
DATE
Setting for the date
TIME
Setting for the time
PASSWORD
4-digits with numbers 1-4
Password protection on activation
SW-VERSION
Only for information
Version of the instrument software
SW-CO.PROC.
Only for information
Version of the coprocessor software
Содержание Filtrax
Страница 2: ......
Страница 4: ...2 Contents Section 8 Spare parts 51 Section 9 Warranty and liability 53 Section 10 Contact 55 ...
Страница 6: ...4 Technical data ...
Страница 12: ...10 General information Fig 2 6 Filter module ...
Страница 16: ...14 Installation Fig 3 9 Module carrier and filter module ...
Страница 31: ...Installation 29 ...
Страница 33: ...Commissioning 31 Fig 4 23 Inserting filter modules ...
Страница 35: ...Commissioning 33 ...
Страница 43: ...Operation 41 ...
Страница 55: ...52 ...
Страница 57: ...54 Warranty and liability ...