![Haag-Streit LENSTAR LS 900 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/haag-streit/lenstar-ls-900/lenstar-ls-900_installation-manual_575789006.webp)
LENSTAR LS 900 / 900 APS
6
© HAAG-STREIT AG, 3098 Koeniz, Switzerland – HS-Doc. No. 7220553 - Edition 3.10 / 2016-02
2.3
Mounting the roller rails/slide plate and cable connection
NOTE!
LS 900 manual cross slide
covered in chapters 2.3.1 – 2.3.3
LS 900 APS automated positioning system
covered in chapters 2.3.4 – 2.3.6
NOTE!
Clean the table surface before positioning the side identification sticker (LS 900 1021635 / LS 900
APS 1022151). Make sure that no oil and grease residue remains on the surface.
NOTE!
Function check and right/left recognition test must be performed after every
new installation
changes on the rail
changes of side identification sticker.
2.3.1
Roller rails/slide plate for LENSTAR LS 900 (not for LS 900 APS)
The roller rails/slide plate must be positioned in such a way, that when the device is centred, the patient’s eye will be
at a distance of 78 mm from the front of the device.
1.) Head rest columns
2.) Rail (1021435)
3.) Aperture for cables
4.) Side identification sticker* (1021635)
5.) Slide plate (1008386)
*
The side identification sticker features a mounting aid to align it properly with the table edge if used in combination with a HS head rest Part
Number: 7200123
1
3
4
5
2