LENSTAR LS 900 / 900 APS
© HAAG-STREIT AG, 3098 Koeniz, Switzerland – HS-Doc. No. 7220553 - Edition 3.10 / 2016-02
9
2.3.4
Roller rails and slide plate for LENSTAR LS 900 APS
The roller rails/slide plate must be positioned in such a way that when the device is centred, the patient’s eye will be
at a distance of 78 mm from the front of the device.
NOTE!
Set of table fittings for LS 900 APS (1022328) include:
2x Rails long (1021435)
2x Countersunk screw 4x20 (3000464)
2x Countersunk screw 4x12 (3000204)
2x Arrestor (1022317)
1x Rail cover complete (1021562)
1x Rail cover (1021556)
1x Magnetic slide plate (1022332)
1x Side identification sticker (1022151)
NOTE!
Clean the table surface before positioning the side identification sticker (1022151). Make sure that no
oil and grease residue remains on the surface.
WARNING!
The aperture of the cables has to be cut free form the side identification sticker with a sharp knife.
Ensure that there are not white or reflecting edges when doing so.
1.) Rail (1021435)
2.) Aperture for cables
3.) Side identification sticker* (1022151)
4.) Slide plate (1022332)
5.) Arrestor (1022317)
6.) Warning sticker “Do not use height adjustment without instrument in place”
*
The side identification sticker features a mounting aid to align it properly with the table edge if used in combination with a HS head rest Part
Number: 7200123.
1
3
2
4
5
6
2