![H-Tec Education E208 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/h-tec/education-e208/education-e208_operating-instructions-manual_3073291015.webp)
www.myhtec.com
15
Technical Data
H x W x D:
Weight:
H
2
Output:
O
2
Output:
Input Voltage Range:
Input Current Range:
Electrode Area:
Guide value for Distilled Water:
Permitted operating pressure:
100 x 80 x 90 mm
0.2 kg
35 mL/min
17.5 mL/min
0 - 2 V
0 - 5 A
*
16 cm
2
<2 µS/cm
0 - 20 mbar
Troubleshooting
The most frequent causes of faults during operation of electrolyzer are:
Use of contaminated water
Use of contaminated water
This electrolyzer cell has a special membrane and it requires pure de-ionized water or distilled
water for its operation. Use of any other water sources (tap water, lake water, sea water,
etc.) will compromise the functionality of the membrane and lower its ionic conductivity. A
contaminated membrane will have significantly lower performance compared to a non-
contaminated membrane.
Reversed polarity connection
Reversed polarity connection
If the power supply connection to the electrolyzer is done wrongly and operated, the cathode
catalyst and cathode gas diffusion layer will be permanently damaged.
Reversed
Reversed membrane after disassembly
membrane after disassembly
If the catalyzed membrane is not placed correctly after the disassembly of the hardware (such as
anode catalyst side placed toward the cathode’s perforated electrode), the cell will not operate or
internal components may get damaged.
Insufficient
Insufficient liquid de-ionized water or running the cell dry
liquid de-ionized water or running the cell dry
If the anode chamber does not get sufficient liquid de-ionized water or distilled water or if it
operates in dry mode, components of the cell will get damaged.
*: The amount of electrical current that can be applied to the electrolyzer cell will depend on the temperature of the
*: The amount of electrical current that can be applied to the electrolyzer cell will depend on the temperature of the
pure water. With room temperature, it is possible to achieve 0 A to 4 A. With slightly warmed water (about 40
pure water. With room temperature, it is possible to achieve 0 A to 4 A. With slightly warmed water (about 40
°°
C), it is
C), it is
possible to achieve 0-5 Amps.
possible to achieve 0-5 Amps. CAUTION: The maximum voltage that can be applied to this cell is 2.1 V.
CAUTION: The maximum voltage that can be applied to this cell is 2.1 V.
Disposal
Do not dispose of electrolyzer cells or components as general household waste.
According to European regulations, used electric and electronic devices may no longer be
disposed of as unsorted household waste. The symbol of the crossed-out wheelie bin indicates
the requirement for separate disposal.
Your local waste management company can provide you with additional information about
disposal options.