GLN-5040A User Manual
12
Set up
Procedure
1.
Ground the GLN-5040A with the testing
receiver ground wire.
2.
Connect the isolated transformer and the GLN-
5040A two line V-network.
3.
Connect to the EUT; please place the equipment
with the reference to the above schematic
diagram.
4.
Connect the GLN-5040A interference signal
output terminal to the EMI receiver input
terminal.
5.
Select the correct filter for the testing standard.
6.
Connect the GLN-5040A to the isolation
transformer, which is connected to the power
grid, and power on to test.
7.
Use the phase line switch on the GLN-5040A to
select different phase lines during the test.
Note
When using the equipment, don't open the case,
don't in a humid, explosive or inflammable
environment. Please ensure the equipment
surface is dry and clean before use.
The maximum voltage should be under 50V
when using the AC power supply.
Please read the user manual carefully before
use to become familiar with the safety the
requirements and to test correctly.
It is recommended to use an isolation