IV.

 

 

STEPS FOR SETTING CURRENT LIMIT 

(1)

 

  Determine the maximum safe current for the device to be powered. 

(2)

 

  Temporarily  short  the  (+)  and  (-)  terminals  of  the  power  supply  together 

with the test lead (

For safety assurance, please select the adequate output 

test lead according to Table 1

).   

(3)

 

  Rotate  the  COARSE  VOLTAGE  control  away  from  zero  sufficiently  to 

have the CC indicator lightened. 

(4)

 

  Adjust  the  CURRENT  control  for  the  desired  current  limit.  Read  the 

current value on the Ammeter. 

(5)

 

  The  current  limit  (overload  protection)  has  now  been  preset.  Do  not 

change the CURRENT control setting after this step. 

(6)

 

  Remove  the  short  between  the  (+)  and  (-)  terminals.  The  power  supply 

returns to the constant voltage mode and is ready to use.   

Table 1. Output Test Lead selection 

Conductor 

Maximum 

Conductive 

Resistor (

/km) 

Permissible 

Current A(amp) 

Wire No. 

AWG 

Cross section Area 

(mm

2

Outer Diameter 

(mm) 

24 

0.22 

0.64 

88.6 

7.64 

22 

0.34 

0.78 

62.5 

10.0 

20 

0.53 

0.95 

39.5 

13.1 

18 

087 

1.21 

24.4 

17.2 

16 

1.32 

1.53 

15.6 

22.6 

14 

2.08 

2.03 

9.90 

30.4 

12 

3.29 

2.35 

6.24 

40.6 

10 

5.23 

3.00 

3.90 

55.3 

 