SWEEP Function
Background
This function can be used in practical applications
that need a waveform with different output voltages.
You can edit the output waveform according to your
needs. The amplitude for the output waveform has a
range equal to the output voltage of the power
supply. The setting range for the output waveform
duration is 0.005S ~ 3600S, with a resolution of
0.001S.
Note: The maximum current is 3.2A when this feature
is active.
Parameter
description
NCycle
This parameter is used for setting the
number of waveform cycles. “0” represents
an infinite cycle. “1” represents one cycle,
“2” represents two cycles, and so on. The
range for this parameter is from 0 to 9999.
Steps
This parameter is used for setting the
number of steps. The range for this
parameter is from 1 to 20.
Direction
This parameter is used for setting the
sweep direction. Single direction (Single)
and double-direction (Double) options are
available.
Curr
The parameter is used to set the current
limit. The range for this parameter is from
0.0000A to 3.2000A.
Parameter
settings
NCycle
In the SWEEP interface, press the F1
function key, enter the number of cycles to
run and then press the
◼ ⧫
⚫
◼ ⧫
key to confirm.
Steps
In the SWEEP interface, press the F2
function, enter the number of steps and