OPERATION
143
Note
Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.
Editing CP Parameters
Background
Constant Power mode can be set to have a
maximum limit (Curr Limit). Using the current
limit enables limiting the current draw.
When using CP mode on a single channel load
module, two power levels can be configured, A
Value and B Value. On a dual channel load
module, only one power level can be configured
per channel: A Value.
Ensure the menu is in CP Mode. See page 138.
Configure
Configure
Mode
CP
Range
Low
Parameters
A Value
0 ~ Setting Max W
B Value
0 ~ Setting Max W
Curr Limit Load module dependent
Panel Operation
1.
Use the Selector knob to highlight
A Value.
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...