![GW Instek PEL-2000B Series Скачать руководство пользователя страница 138](http://html1.mh-extra.com/html/gw-instek/pel-2000b-series/pel-2000b-series_user-manual_2280092138.webp)
PEL-2000B Series User Manual
138
Select CV Mode
Background
The PEL-2000B series electronic load operates in
four different modes, Constant Current (CC),
Constant Resistance (CR), Constant Voltage (CV),
and Constant Power (CP).
CV Mode cannot be used with the Group Unit
mode.
When a channel is active, the F1 key can
be used to switch between each
operating mode.
F1
F5
~
F1
Mode
CC
Mode
CR
Mode
CV
Mode
CP
Panel Operation
1.
Press the CHAN button and use
the selector knob to select a
channel.
CHAN
LOCAL
2.
Press F1 until CV mode is
displayed in the display panel.
F1
F5
~
F1
Configure
Configure
Mode
CV
Response
Slow1
Range
High
Note
Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...