LCR-914/915/916 User Manual
22
Shock Vibration:
Sinusoidal vibration per MIL-T- 28800E
(5 ~ 55 Hz, 3g maximum).
Drop Protection:
4 feet drop to hardwood on concrete floor.
Indoor Use.
Electrical Specifications
(1) Test Frequency
Range
Resolution
Accuracy
100.00 Hz
0.01 Hz
± 0.01%
120.00 Hz
0.01 Hz
± 0.01%
1.0000 kHz
0.1 Hz
± 0.01%
10.000 kHz
[1]
1 Hz
± 0.01%
100.00 kHz
[2]
10 Hz
± 0.01%
[1] The 10kHz test frequency is only for the LCR-915/916.
[2] The 100kHz test frequency is only for the LCR-916.
(2) Test Signal
AC Signal Level:
600mVrms
AC Signal Accuracy:
±10%
DC Bias Level:
1V
DC Bias Accuracy:
±10%
(3) Test Cable
Model
Length
Bandwidth
Type
SMD Test Probe
60cm
1MHz
5-Wire
4-Wire Test Probe
60cm
1MHz
5-Wire
Alligator Clip Set
15cm
1kHz
2-Wire
Accuracy:
±(A x B)(% of reading)
A: Basic Accuracy as specified by
B: Test Cable Accuracy
B(%) = 1 + (L x F x T)
L(m): Cable length
F(MHz): Test frequency
T: Cable type. If the cable is 5-
wire type, then “T” is 40, else “T”
is 4,000.
When measuring by basic accuracy, the following
conditions must be met:
1.
Ambient temperature: 23˚C ± 5˚C < 80% RH.
2.
Test cable length: 0m
3.
Open and short calibration has been performed
4.
D ≤ 0.1 for C, L measurements; Q ≤ 0.1 for R
measurements.
See the operation manual for additional details.