LCR-6000 Series User Manual
31
Table 4-12
LCR-6002’s predefined test frequencies that can be
selected by using INCR+/DECR
‐
INCR+/DECR-
10Hz
50Hz
60Hz
100Hz
120Hz
1kHz
2kHz
4.1.4 Trigger Mode [TRIG]
LCR-6000 Series supports four trigger modes: INT
(internal), EXT (external), MAN (manual) and BUS (RS-
232).
Trigger
Mode
Description
INT
LCR-6000 Series continuously repeats the
measurement cycle.
MAN
LCR-6000 Series performs one cycle of
measurement each time you press the [Trig]
key.
EXT
LCR-6000 Series performs one cycle of
measurement each time a rising pulse is input
to the handler external trigger input pin on
the rear panel.
BUS
LCR-6000 Series performs one cycle of
measurement each time it receives a trigger
command sent via RS-232.
Procedure for choosing trigger mode [TRIG]
Step 1. Press the [Measure] key.
Step 2. Use the cursor key to select the [TRIG] field.
Step 3. Use the soft-keys to select desired trigger
mode.
Soft-key
Function
INT
Internal Trigger Mode
MAN
Manual Trigger Mode
EXT
External Trigger Mode
BUS
BUS Trigger Mode