GDP Series User Manual
14
Caution
The power light must come on
Ensure the range light does not come on before
measuring
Deviation Table
Model
GDP-050
GDP-100
GDP-050
GDP-100
All
GDP-050
GDP-100
GDP-025
GDP-025
Attenuation
X 1000
X 500
X 200
X 100
X 50
X 20
Voltage input range
(DC + AC Peak)
7000 Vp-p
(±3500Vdc)
3500 Vp-p
(±1750Vdc)
1400 Vp-p
(±700Vdc)
700 Vp-p
(±350Vdc)
350 Vp-p
(±175Vdc)
140 Vp-p
(±70Vdc)
Vertical Deviation
on the
Oscilloscope in
V/div
Real Division in V/div
X 1000
X 500
X 200
X 100
X 50
X 20
Model
GDP-050
GDP-100
GDP-050
GDP-100
All
GDP-050
GDP-100
GDP-025
GDP-025
1
1000
500
200
100
50
20
0.5
500
250
100
50
25
10
0.2
200
100
40
20
10
4
0.1
100
50
20
10
5
2
50m
50
25
10
5
2.5
1
20m
20
10
4
2
1
0.4
10m
10
5
2
1
.5
0.2
5m
5
2.5
1
0.5
0.25
0.1
2m
2
1
0.4
0.2
0.1
40m
Note
The real vertical deviation in V/div is equal to the
attenuation factor multiplied by the range of the
vertical deviation selected on the oscilloscope. The
real vertical deviation will be doubled if a 50
Ω
load is
used.
Содержание GDP-025 Series
Страница 21: ...Current Probe and Power Supply USER MANUAL GW INSTEK PART NO 82CP 425P0M01 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...TABLE OF CONTENTS 4 ...
Страница 34: ...GCP Series User Manual 14 Names of Parts Current Probe External view ...
Страница 38: ...GCP Series User Manual 18 Front view GCP 206P Rear View GCP 206P ...
Страница 39: ...OVERVIEW 19 Power Supply Receptacle ...
Страница 41: ...OVERVIEW 21 4 Turn the GCP 206P or GCP 425P power switch ...