background image

  GDM-8246 MULTIMETER 

PROGRAMMER MANUAL 

 

  46  

 

7.    STATUS AND ERROR REPORTING 

 

A set of status registers allows the user to quickly determine the 

DMM’s internal processing status. The status register, as well as the 
status and event reporting system, adhere to SCPI recommendations. 

Structure of System 

The sketch of the status and event reporting system is showed as   
figure 7. Each component of the sketch represents a set of registers and 
queues that can read, report, or enable the occurrence of certain events 
within the system. 

If a specific event in the DMM sets a bit in a 

status register

, reading 

which can tell you what types of events have occurred. 

Each bit in the status register corresponds to a bit in an 

enable register; 

the enable bit must be high for the event to be reported to the Status Byte 
Register. 

A Service Request (SRQ) is the last event to occur. The SRQ requests an 
interrupt on the GPIB to report events to the system controller. 

Status Registers 

There are two kinds of status registers are included to the DMM. 

z

 

OPERation Status Registers ( CONDition, EVENt, and ENABle) 

z

 

QUEStionable Status Registers (CONDition, EVENt, and ENABle) 

 

The lower level nodes: QUEStionable and OPERation each have three 16 
bits registers: CONDition, EVENt, and ENABle. Figure 8 shows the 
sequential relationship between these three types of registers and the 
commands that relate to each register. 

 

  GDM-8246 MULTIMETER 

PROGRAMMER MANUAL 

 

  47  

 

Frequency Null Sense

Not  Used

Not  Used

Limit Test Fail HI

Limit Test Fail LO

Capacitance Overload

Ohm Overload

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

Voltage Overload

Current Overload

QUEStionable Status

0

1

2

3

9

15

14

13

12

11

10

8

7

6

5

4

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

Not  Used

OPERation Status

Not  Used

2

3

0

1

14

13

12

11

10

9

8

7

6

5

4

15

Command Error

Power On

User

Request

Execution Error

Device  Dependent

Error

Query Error

Not Used

Standard Event Status Registers

Operation Complete

ESB

RQS/MSS

MA

V

Not Used

Status Byte Register

Not Used

Summary of IEEE 488.2 Status Structure Registers

2

3

0

1

7

6

5

4

2

3

0

1

7

6

5

4

E/E

Not  Used

Not  Used
Not  Used
Not  Used
Not  Used
Not  Used
Not  Used

QUES

OPER

SRQ

Error/Event Queue

Output Queue

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 7. A graphic representation of the status registers and their connections. 

 

Содержание GDM-8246GP

Страница 1: ...The measured procedures can be varied with user s testing programs therefore the Digital Multimeter can be operated remotely from an instrument controller or computer across the RS232 interface option...

Страница 2: ...ace function DC1 Device Clear The DMM has Device clear capability to return the device to power on status DT0 Device Trigger The DMM has no Device Trigger interface function C0 Controller The DMM can...

Страница 3: ...e both sides Once you have set these parameters you can control the DMM through the RS232 interface z Baud rate You can set rates of 1200 2400 4800 or 9600 baud z Parity bit none z Data bit 8 bits z S...

Страница 4: ...nection 9 No connection Figure 1 Pin assignments of the RS232 connector on the rear panel for DB 9 D GDM 8246 MULTIMETER PROGRAMMER MANUAL 7 Figure 2 Wiring configuration for DB9 to DB9 Computer s Con...

Страница 5: ...equipment manufacturers The IEEE 488 2 syntax has been adopted by SCPI to provide common commands for the identical functions of different programmable instruments GDM 8246 MULTIMETER PROGRAMMER MANUA...

Страница 6: ...oot level A root node and one or more lower level nodes form a header path to the last node called the leaf node Figure 4 Tree hierarchy Leaf SYSTem AUTO Root node ERRor STATe STARt CYCLe Lower level...

Страница 7: ...t must add a LF to be a judgment for message terminator As for query command the return message of the instrument is also added a LF for PC to judge message terminator Entering Commands The standards...

Страница 8: ...nds into four functional classifications z General Setting Commands z Calculating Commands z Status Commands z Miscellaneous Commands The tables also provide a brief explanation of each command GDM 82...

Страница 9: ...LIMit STATe Returns Compare mode on or off CALCulation LIMit STATe Boolean Sets Compare mode on or off CALCulation LIMit LOWer Returns the value of the lower limit CALCulation LIMit LOWer NR2 Sets the...

Страница 10: ...us Register SESR GDM 8246 MULTIMETER PROGRAMMER MANUAL 19 SRE NR1 Sets contents of Service Request Enable Register SRER SRE Returns contents of Service Request Enable Register SRER STB Reads Status By...

Страница 11: ...ation of commands by design SYSTem ERRor Read the next item from the error event queue SYSTem VERSion Returns the SCPI version level GDM 8246 MULTIMETER PROGRAMMER MANUAL 21 6 DETAILS OF COMMAND REFER...

Страница 12: ...al number and firmware version GDM 8246 MULTIMETER PROGRAMMER MANUAL 23 Examples IDN Returns GW_Inc GDM 8246 FW1 00 OPC Function The command form OPC sets the operation complete bit bit 0 in the Stand...

Страница 13: ...contains the binary value 0101 0001 WAI no query form Function WAI prevents the programming instrument from executing further commands or queries until all pending operations are finished Syntax WAI G...

Страница 14: ...gure MODe might return 1 to indicate the MIN mode to be slected Please refer to the representative total value for every attached calculation mode as follows GDM 8246 MULTIMETER PROGRAMMER MANUAL 27 T...

Страница 15: ...the continuity function Syntax CONFigure CONTinity GDM 8246 MULTIMETER PROGRAMMER MANUAL 29 Examples CONFigure CONTinity sets the continuity function Please refer to the Table 6 1 CONFigure CURRent DC...

Страница 16: ...lease refer to the unit for every range shown as Table 6 1 CONFigure SFRequency Function GDM 8246 MULTIMETER PROGRAMMER MANUAL 31 Set the frequency function in the AC mode Syntax CONFigure SFRequency...

Страница 17: ...e VOLTage ACDC NR2 Function Set the ACDC voltage function and range Syntax CONFigure VOLTage ACDC NR2 Examples CONFigure VOLTage ACDC 0 sets the AC DC voltage function and auto range CONFigure VOLTage...

Страница 18: ...voltage SVALue query only Function Returns the value displayed on the primary and secondary display with 6 characters totally GDM 8246 MULTIMETER PROGRAMMER MANUAL 35 Syntax SVALue Examples SVALue mi...

Страница 19: ...function CALCulation LIMit UPPer 1 0000 sets 1 0000 DC voltage at the upper limit in the DC voltage function CALCulation LIMit FAIL Function Returns limit result Syntax GDM 8246 MULTIMETER PROGRAMMER...

Страница 20: ...mode Syntax CALCulation RELation DATa CALCulation RELation DATa NR2 GDM 8246 MULTIMETER PROGRAMMER MANUAL 39 Examples CALCulation RELation DATa might return 1 0000 in DC voltage function CALCulation R...

Страница 21: ...hold mode GDM 8246 MULTIMETER PROGRAMMER MANUAL 41 CALCulation HOLD 1 sets the Hold mode STATus OPERation CONDition query only Function Return the contents of the OPERation register The DMM however do...

Страница 22: ...n destructive GDM 8246 MULTIMETER PROGRAMMER MANUAL 43 Syntax STATus QUEStionable CONDition Returns NR1 Examples STATus QUEStionable CONDition returns 0 STATus QUEStionable ENABle Function Set or quer...

Страница 23: ...le EVENt Returns NR1 Examples STATus QUEStionable EVENt returns 0 SYSTem ERRor query only Function Query the next error message from the Error Event queue The result of the query is the error number f...

Страница 24: ...Registers CONDition EVENt and ENABle z QUEStionable Status Registers CONDition EVENt and ENABle The lower level nodes QUEStionable and OPERation each have three 16 bits registers CONDition EVENt and...

Страница 25: ...Table 4 shows the bit designations of the 16 bit QUEStionable Status Register Enable Register Event Register Condition Register To SBR GDM 8246 MULTIMETER PROGRAMMER MANUAL 49 Table 4 QUEStionable St...

Страница 26: ...gister GDM 8246 MULTIMETER PROGRAMMER MANUAL 51 Use the serial poll or the STB Query to read the contents of the SBR The bits in the SBR are set and cleared depending on the contents of the Standard E...

Страница 27: ...hich types of events are summarized by the Event Status Bit ESB in the SBR The bits of the ESER correspond to the bits of the SESR Use the ESE command to set the bits in ESER Use the ESE query to read...

Страница 28: ...put queue stores the message The output queue stores and reports the messages on a FIFO first in first out state The SYSTem ERRor query reads the next item from the output queue If output queue overfl...

Отзывы: