Option Mounting Methods: Either Base mounting or VESA mounting can be used.
"Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards (e.g. IEC 60950 for data processing equipment
and IEC 60601-1 for medical equipment). Furthermore all configurations shall comply with
the valid version of the system standard IEC 60601-1-1. Everybody who connects additional
equipment to the signal input part or signal output part configures a medical system, and is
therefore responsible that the system complies with the requirements of the valid version of
the system standard IEC 60601-1-1. If in doubt, consult the technical service department or
your local representative. " (MDD: 13.6.c, IEC 60601-1: 6.8.2.c, 19.2.b, 19.2.c)
External Equipment
External equipment intended for connection to signal input/output or other connectors, shall
comply with relevant UL/IEC standard (e.g. UL 1950 for IT equipment and UL 60601-1 / IEC
60601 series for systems – shall comply with the standard IEC 60601-1-1, Safety
requirements for medical electrical systems. Equipment not complying with IEC60601-1-1
shall be kept outside the patient environment, as defined in the standard.