19
3. Check the electrical condition of the products
• Check the batteries and measure the:
• Input/Voltage
• Output/Voltage
• Check electrical functions and signals
• Check all wirings for errors and defects
• Check wire lead-ins
• Check all possible conections, plugs etc.
4. Check the mechanical condition oft the product
• Clean the product for dirt and other impurities
• Inspect and evaluate the vital parts of the product
• Exchange defective and worn parts
• Grease the product
5. Point 1.2 is carried out one more time as a control function
6. Have new errors or problems arisen in point 7?
• If new problems have arisen, go back to point 3
• If no problems have arisen, finish the inspection
5.04
Trouble Shooting
If the LP12 does not react when pressing the keys of the control panel?
• Check that the emergency stop is not activated
• Check that the batterybox is properly connected
• Please contact your local Guldmann representative
If the green light does not come on when pressing the keys
of the hand control?
• Check that the emergency stop is not activated
• Check that the batteries are charged
• Check if the hand control is faulty or if the LP12 has a blown fuse
• Please contact your local Guldmann representative
If the red or yellow light is flashing after the LP12 has been recharged?
• Check that the light on the charger is on when recharging
• Check that the charger is on
• Check that the charging plug have been connected properly to the charging
station
• Please contact your local Guldmann representative