6
© Guldmann GB/US-1706/12/2016 • # 550865_14
© Guldmann GB/US-1706/12/2016 • # 550865_14
Button for reset of emergency stop
GH1 is delivered with the emergency stop activated in order to ensure that the
battery is not being discharged during long-term storage. Reset the emergency
stop by pressing the “RESET” button on the bottom of the hoist
(see point 2.04)
.
Mounting of side cover
GH1 is standard delivered with side cover in
white and yellow.
Before starting to mount the side cover it is
important, for safety reasons, to make sure
that the emergency stop is activated
(see point 2.04).
The chosen side covers are fitted on the side
of GH1 by bending the cover slightly and
placing it in the groove at the side of GH1.
1.07
Installing a new GH1 hoist in an existing rail system
Please notice, when installing a new GH1 hoist in an existing rail system it
must be ensured that:
• The rated max load of the rail system, must be equal or higher than the max
load of the new hoist.
– If there is no max load mentioned on the rail system, the rail system must
then be checked according to the guideline in the installation manual
(distance between bracket according to max load).
– If the brackets are not visible, then a load test with 1,5 x max load of the
hoist must be performed minimum 20 min. During the test the deflection of
the rails must not be higher than 1/200 of the length of the rail.
– If it is not possible to do any of the above mentioned, please contact
Guldmann or their representative.
• If the rail system can not be rated to the same max load as the hoist, then
extra brackets must be installed according to the installation manual (distance
between bracket according to max load).
Содержание GH1