Section 1
TM6564-C-00
26 April, 2011
1-2
1.2.
SPECIFICATIONS
NUMBER OF INPUTS
8 or 16 for Model 6564
ISOLATION **
700 Peta-Ohms Typical, 100 Peta-Ohms Minimum
SYSTEMATIC ERROR DUE CHANNEL VARIABILITY **
Leakage will vary by 3.3% of 100 Peta-Ohms typical, 5% of 100 Peta-Ohms
maximum.
ENVIRONMENTAL LIMITS
Operating:
10
o
C to 40
o
C up to 80% relative humidity
Storage:
-20
o
C to 65
o
C up to 95% relative humidity
** Note: Specifications apply only if temperature is stable within 1
o
C, free of
drafts and the relative humidity is below 70%. Warm the scanner up for 2 hours
min.
SCANNER CONTROL
Local, using front panel push-buttons
Remote, via IEEE-488 bus (interface included)
RELAY CONTACT RATINGS
Life: greater than 10,000,000 cycles at low levels
Current: 2 amps maximum at 10 volts
Voltage switched: 100 volts maximum at 1 milliamp
Voltage non-switched*: 1000 volts max. (for terminal inputs)
*CAUTION - reduce voltage before actuating relays.
NOTE: when applied voltage is above 30 volts, the current source must limit the
current to less than 2 milliamps to meet IEC 1010-1/EN61010-1 safety
requirements.
SIZE
Length: 420 mm (16.5 in.)
Width: 451 mm (17.7 in.)
Height:133 mm (5.2 in.)
WEIGHT
16 channel scanner: 10 kg (23 lb.)
LINE POWER
100V, 115V-127V, 220V-230V, 240V all
±
10%; 50-60 Hz