Guardian Voyager Portable Ceiling Lift System 98000 Скачать руководство пользователя страница 6

Voyager Portable

M a i n t e n a n c e   &   C a r e

11

Voyager Portable

10

M a i n t e n a n c e   &   C a r e

Strap Inspection

This equipment is built with a strap that can withstand a load of 3000 lbs. (1364 kg).
However, if this strap is damaged or shows signs of wear, the acceptable load on the strap
before rupture can drop rapidly and present a danger for the person being lifted or the
caregiver.

Sunrise Medical recommends thoroughly inspecting the strap every two months as follows:

1. Completely unwind the strap.
2. Look for any of the signs of wear described in (Figure 9).

If there is any sign of wear as indicated here or other visual defect, the strap
should be changed immediately. By continuing to use the lift without changing
the strap, caregiver and user safety is greatly compromised. In any case, Sunrise
recommends to change strap at least every two years. 

NOTE– 

Make note of results of inspection in the logbook on the back of this manual.

HANDLING AND STORAGE

Avoid violent shock during transportation.

The lift should not remain stored for long periods of time without recharging the batteries.

NOTE–

Sunrise Medical recommends charging of batteries at least every two weeks even if the lift is
not used. This will prevent premature aging of batteries.

BATTERY REPLACEMENT

Sunrise uses sealed lead-acid batteries. Unlike nickel-cadmium, these batteries do not have
any memory effect. Therefore, batteries should not be completely drained before recharge.

Replace the battery when it has been dropped or if there is a noticeable reduction in the
number of transfers that can be performed between charges. If you notice the Voyager
Portable “beeping”, see the instructions in “Troubleshooting” to ensure it is a problem with
the battery.

To replace a battery, contact your local provider and order product number 98850.

NOTE– 

DO NOT ATTEMPT TO USE A BATTERY NOT SPECIFIED BY SUNRISE MEDICAL.

Guardian batteries are specially designed for Guardian charging systems.
Attempting to use an unauthorized battery may seriously damage the lift and/or
the charger or cause serious injury.

Loose threads in stitched area.

Noticable discoloration 

by

having a  lighter  color  strap  than
the double 

thickness  stitched

area.

Side wear

Middle wear

Figure 9

Loose threads in
stitched area

Noticable discol-
oration– strap is
lighter in color than
the double thickness
stitched area.

Side wear

Middle wear

MAINTENANCE & CARE

Always carry out the daily checklist before using the lift.

Unauthorized alterations made to the Voyager Portable could cause serious injury.

The Voyager Portable and accessories should be thoroughly inspected periodically.
Sunrise Medical recommends an authorized service dealer perform the inspec-
tions. Daily checks should be performed by the user before use.

Preventative maintenance specified in this manual can prevent accidents and
reduce repair costs.

DAILY CHECKLIST

The following procedures must be followed before each use:

Charge the battery whenever the lift is not in use.

Inspect the battery before placing it in the Voyager Portable. If the battery appears
to have any damage, DO NOT USE. Replace the battery immediately.

Inspect the lift for any damage. If the lift casing does not look properly aligned, if
there are any cracks or other damage on the lift or if any parts are missing, DO NOT
USE. Contact your local provider to have the lift serviced.

Attach the lift to the trolley in the rail. Inspect the visible strap for any signs of
wear, frays, loose threads or other damage. If there is any evidence of damage, DO
NOT USE. Contact your local provider to have the lift serviced.

Inspect the sling for tears, frayed straps or loose stitching. If the sling has any of
the above damage, DO NOT USE. Contact your local provider to have the sling
replaced.

Inspect the reacher and adjustable extension strap (if you have these options) for
wear. Check the adjustable extension strap for frays or loose stitching. Check the
reacher for damage in the metal hook or handle. DO NOT USE if there is any damage.

Inspect the hook at the top of the strap to ensure it is properly attached and there is
no damage to the hook.

Make sure the lift is free to traverse the rail.

Check the control buttons for correct functioning.

INSPECTION AND CLEANING

Clean the Voyager body with a soft dry cloth or a soft cloth lightly moistened with a mild
detergent solution. Do not use any type of solvent that may damage the finish.

DO NOT immerse lift in water.

To ensure a better rolling surface for the trolley wheels, clean the inside of the track every
four months. To do so, insert a damp cloth in the opening and slide it from one end of the
track to the other. 

Always replace rail end stopper (if they have been removed) after servicing.

Содержание Voyager Portable Ceiling Lift System 98000

Страница 1: ...Installation Instruction Manual Guardian Voyager Portable Ceiling Lift System Model 98000...

Страница 2: ...rol and may compromise safety DO NOT store the charger in a shower bath or other area with high humidity DO NOT drop the lift or battery Dropping the battery or lift may cause internal damage that is...

Страница 3: ...ls with adhesive tape etc oth erwise the residual electricity in used batteries may cause fire or explosion EQUIPMENT WARNING LABELS Inspect all precautionary labels on the equipment Order and replace...

Страница 4: ...e low battery buzzer sounds be sure to recharge the battery as soon as possible O p e r a t i o n 7 OPERATION Read Safety Instructions and Warnings before attempting to use the Voyager Lift Always per...

Страница 5: ...Very humid Moving or vibrating CAUTION Do not attempt to put a damaged battery into a charger Inspect the battery first for dents or damage Charging the battery with charger 98820 1 Eject the battery...

Страница 6: ...area Noticable discol oration strap is lighter in color than the double thickness stitched area Side wear Middle wear MAINTENANCE CARE Always carry out the daily checklist before using the lift Unaut...

Страница 7: ...be inside the charger 2 Check to make sure the power cord is correctly plugged in to the charger and the wall 3 If the light still doesn t work plug the charger into a different outlet in a different...

Страница 8: ...and charger checked by your local provider 1 Check if there is any damage to the battery that would interfere in the lift battery s slot IF THERE IS DAMAGE DO NOT USE THE BATTERY 2 Try to slide in ano...

Страница 9: ...Available in sizes Small 90010 Medium 90011 and Large 90012 Guardian Quick fit Sling Figure 14 The Quick fit Sling is specially designed for persons who are stocky or obese especially in the midsecti...

Отзывы: