Operating Instructions
16
PC software
2.2 Zero point
Zero calibration (1st calibration point), this requires a certain number of
samples. To start, click Enter button, beside the raw value cell. These will be
incremented by triggering “Save” or “Abort” button. Zero calibration should
be performed without any flow.
2.3 Interpolation point
Interpolation point calibration, this requires a certain number of samples.
Click the Enter button, beside the raw value cell. These will be incremented
by triggering “Save” or “Abort” button. Pressing the “Save” button will open
an Input window for the reference value. This can be edited at a later stage.
Calibration point and raw values can be edited at any stage.
PC configuration
To save this calibration values it is necessary to activate the check box
“Overwrite Calibration” on the left side of the window and also click the
“Device program” button. This must be done in order to save the results.