7. Wiring Diagrams
PNEG-2175CE
CE Compliant Sampler - Model GCE 45 and GCE 90
29
CE Control Electrical Layout (Continued)
1. The control requires a 400 Volt, 3 PH and neutral supply (TPN) with minimum current capacity of
10 Amp.
2. Run suitable type cables to the components on the Sampler and Divider as indicated in the
NOTE:
The conductor cross section areas stated above the minimum and should be increased if
required, according to local regulations, length of cable and voltage drop.
3. When making connections to the sampler and divider components, refer to the certified drawing
provided with your controls (See
and
Electrical Connections
Item
Voltage
Maximum Current
Terminal (Block) Numbers
Power Supply (TPN)
400 VAC
10 Amp
(X1) 1, 2, 3, N, PE
Service Disconnect Switch
400 VAC
10 Amp
(X1) 4, 5, 6, 7, 8, 9, PE
Sampler Motor
400 VAC
7 Amp
(X2) 1, 2, 3, PE
Divider Motor
400 VAC
1 Amp
(X2) 4, 5, 6, PE
Sampler Motor Thermal Protection
24 VDC
1 Amp
(X3) 1, 2
Divider Motor Thermal Protection
24 VDC
1 Amp
(X3) 3, 4
Right Limit Switch
24 VDC
1 Amp
(X3) 5 (COM), 6 (N.C.), 11 (N.O.), PE
Left Limit Switch
24 VDC
1 Amp
(X3) 7 (COM), 8 (N.C.), 12 (N.O.), PE
External Emergency Stop
24 VDC
1 Amp
(X3) 9, 10, PE