9.3.3 Blank test

For some parameters an additional blank test is necessary, as the 
reagents have a certain inherent colour, which occurs even if 
there is no co-reactant at all. The zero value described above is 
based upon the characteristics of the system "photometer, vial 
and sample" and therefore is the same for all parameters.
For doing the blank test, the measurement has to be executed 
with the diluting water instead of the sample in order only to 
determine the colouring caused not by the reaction. All analysis 
steps, such as reagent-adding and waiting times, are the same. 
It is also possible to choose either detailed operator guidance or 
expert mode.
After the blank test, no result will be shown. It is, however, stored 
by the instrument and will later be subtracted from the measuring 
result.

See section 

13. Error messages and error correction

.

9.3.4 Dilution

If the measuring result exceeds the measuring range (the error 
message "value exceed" is displayed), the sample water can be 
diluted with distilled water. See section 

6.5 Measuring ranges and 

tolerances

.

In this case, proceed as follows:
• At the step "fill in 10 ml sample" requested on the display, first 

add 9 ml distilled water with the syringe.

• Then add 1 ml sample water with the syringe, close the vial 

and shake.

• Continue with the measurement as requested on the display.
• Multiply the measuring result by 10: result in mg/l or ppm.

9.4 Aluminium

Reagent not for sea-water use!

Before determining the aluminium concentration, a blank 
test has to be made. See section 

9.3.3 Blank test

.

9.4.1 Aluminium determination

1. Select "photom." in the main menu, and press [OK].
2. Select "aluminium", and press [OK].
3. Select "measure" or "blanktest", and press [OK].
4. Fill a standard vial with 10 ml of the water sample. In case of 

blank test: 10 ml of the diluting water. Press [OK].

5. Add one level spoon (grey) of Al reagent 1, close the vial and 

shake. Press [OK].

6. Add three drops of Al reagent 2, close the vial and shake. 

Press [OK].

7. Add four drops of Al reagent 3, close the vial and shake. 

Press [OK].

8. Wait for seven minutes.
9. Insert the vial, start the measurement with [OK], and read the 

result.

– 

Result: aluminium concentration in mg/l or ppm

.

10.Press [OK].

– Printer or memory being selected: Print or save result. 

Measurement is finished.

– Printer or memory not being selected: Measurement is 

finished.

Caution

The blank test is necessary after commissioning, 
after battery exchange or recharging of a totally 
discharged accumulator and before the use of 
new reagent batches.

Caution

Carry out the dilution process very carefully in 
order to guarantee the utmost accuracy.

Determination 
method

Reaction of Al

3+

 with Chromazurol-S in 

weakly acetate-buffered medium with 
production of a red-violet colorant

Measuring range

0.02 to 0.5 mg/l Al

3+

Reagent 
temperature range

Higher than 25 °C

Analysis time

About seven minutes

Number of 
analyses per 
reagent set

About 200

Warning

Reagent 2 contains dimethylsulfoxide.

R 36/38: Irritating to eyes and skin.

S 26: In case of contact with eyes, rinse 
immediately with plenty of water and seek 
medical advice.

Note

At storage under 20 °C the elements of reagent 2 
crystallise. If that is the case, warm up the 
reagent slightly and shake.

Содержание DIT

Страница 1: ...DIT Photometer GRUNDFOS ALLDOS INSTRUCTIONS Installation and operating instructions...

Страница 2: ...States relating to Electromagnetic compatibility 89 336 EEC Standards used EN 61326 1997 A1 1998 A2 2001 Class B EN 61326 1997 A1 1998 A2 2001 EN 60950 12 2001 power pack EN 61000 6 2 08 2002 power pa...

Страница 3: ...ttery exchange DIT N DIT S 26 12 2 Recharging the accumulator DIT S 26 12 3 Cleaning 27 13 Error messages and error correction 27 13 1 Errors and defects of hardware and software 27 13 2 Error message...

Страница 4: ...ce 5 2 Avoidance of danger If safe operation cannot be ensured any longer the DIT Compact Photometer is to be put out of service and safeguarded against unintentional operation This is the case under...

Страница 5: ...language depending on software version Three operating modes can be selected In normal mode all analysis steps will be displayed The expert mode for the proficient user shows shortened texts and does...

Страница 6: ...ence of temperature on the pH value can be compensated and displayed by using electrodes with built in thermocouple Different types of buffer solutions can be used for calibration With all versions th...

Страница 7: ...pH 9 18 pH Grundfos Alldos 4 01 pH 6 86 pH 9 18 pH DIN NIST Measuring range 1190 mV Resolution 1 mV Accuracy 4 mV within 700 mV to 700 mV 7 mV outside of 700 mV to 700 mV Error message for leaving me...

Страница 8: ...rument will be activated and continues at the last position of the program Therefore there is no separate button for switching on or off the instrument 8 2 Operating buttons There are four buttons for...

Страница 9: ...e cancel cancel language time date parameter profi mod cancel other parameters config cal data cancel unit version document measure blanktest zerocheck cancel ALLDOS DIN NIST others cancel C F auto of...

Страница 10: ...me way Confirm with OK after which the instrument returns to the configuration menu 8 4 4 Parameter This menu item is used for taking single parameters out of the menu photom This is useful if not all...

Страница 11: ...will be executed Then the instrument returns to the parameter menu 100 results can be stored When reaching this an error message will appear and shortly after that the message print memory The Up and...

Страница 12: ...ach measuring the zero adjustment can be checked When having selected a parameter it is possible to select zerocheck The instrument asks the user to fill a vial with 10 ml of the sample water and inse...

Страница 13: ...k test has to be made See section 9 3 3 Blank test 9 4 1 Aluminium determination 1 Select photom in the main menu and press OK 2 Select aluminium and press OK 3 Select measure or blanktest and press O...

Страница 14: ...le Measuring range 0 2 to 3 0 mg l NH4 Reagent temperature range 15 C to 25 C Analysis time About 12 minutes Number of analyses per reagent set About 250 Warning Reagent 1 contains sodium hydroxide R...

Страница 15: ...Result concentration of total chlorine in mg l or ppm 15 Press OK Printer or memory being selected Print or save result The results for free total and combined chlorine are shown in the display in mg...

Страница 16: ...t photom in the main menu and press OK 2 Select ClO2 and press OK 3 Select measure and press OK 4 Put three drops of Cl2 reagent 1 into vial Cl2 1 Press OK 5 Add one drop of Cl2 reagent 2 close the vi...

Страница 17: ...Blank test 9 10 1 Chromium determination 1 Select photom in the main menu and press OK 2 Select chromium and press OK 3 Select measure or blanktest and press OK 4 Fill a standard vial with 10 ml of t...

Страница 18: ...asurement and clean it with the brush as the reagent causes a slight opaqueness Determination method After adding CN reagent 1 chlorine is set free which reacts with cyanide ions to chlorine cyane Thi...

Страница 19: ...nter or memory not being selected Measurement is finished 9 Rinse the syringe well after the determination Determination method Iron III ions are reduced to iron II ions The iron II ions react with 2...

Страница 20: ...memory not being selected Measurement is finished Determination method Hydrazine reacts with 4 dimethylamino benzaldehyde in an acid solution to yellow 4 dimethylamino benzaldehydrazine Measuring rang...

Страница 21: ...Wait for five minutes 10 Insert the vial start the measurement with OK and read the result Result nitrate concentration in mg l or ppm 11 Press OK Printer or memory being selected Print or save result...

Страница 22: ...ne concentration in mg l or ppm 9 Press OK Printer or memory being selected Print or save result Measurement is finished Printer or memory not being selected Measurement is finished Determination meth...

Страница 23: ...section 9 3 3 Blank test 9 22 1 Phosphate determination 1 Select photom in the main menu and press OK 2 Select phosphate and press OK 3 Select measure and press OK 4 Fill vial PO4 with 10 ml of the wa...

Страница 24: ...menu item using the Up and Down buttons and confirm by pressing OK Then the DIT returns to the previous menu 10 2 Unit of temperature The menu item C F in the configure menu makes it possible to defin...

Страница 25: ...ll appear first The first digit flashes and the temperature can be adjusted as usual with the Up and Down buttons Confirm with OK after this Now the pH measuring starts The instrument will measure the...

Страница 26: ...and remove the plastic cover Observe the correct polarity of the battery Then refit the cover 12 2 Recharging the accumulator DIT S If the accumulator voltage drops below 7 1 V the following message w...

Страница 27: ...sive cleaning detergents Do not scratch the glass sides in the vial hole when cleaning them The vial hole may only be cleaned with a dry and soft cloth Error message Cause Correction 1 printer error a...

Страница 28: ...28...

Страница 29: ...29...

Страница 30: ...30...

Страница 31: ...n Bldg 5F 5 21 15 Higashi gotanda Shiagawa ku Tokyo 141 0022 Japan Phone 81 35 448 1391 Telefax 81 35 448 9619 Korea GRUNDFOS Pumps Korea Ltd 6th Floor Aju Building 679 5 Yeoksam dong Kangnam ku 135 9...

Страница 32: ...www grundfosalldos com Being responsible is our foundation Thinking ahead makes it possible Innovation is the essence 15 730057 V6 0 Repl 15 730057 V5 0 GB 96681481 0808...

Отзывы: