14. Maintenance

14.1 Intervals for cleaning and maintenance

• Clean the filter

– if high contamination is evident, or if the pressure drops 

considerably.

• Clean the whole flow armature

– if high (visible) contamination is evident.
– if faults occur.
– before starting up again after a prolonged stoppage.

• Replace the electrodes.

– Grundfos recommends that you replace the reference 

electrode and, if necessary, the redox or pH single-rod 
electrode after 12 months.

– The measuring electrode is wear-resistant and does not 

have to be replaced on a regular basis. 

14.2 Cleaning and replacing the filter

• Required tools and aids:

– water
– soft brush.

1. Close the connection to the sample water inlet (U1) and 

outlet (U2).

2. Unscrew the filter cartridge (P).
3. Unscrew the plastic screw thread (P1) at the upper end of the 

filter cartridge (P), using a screwdriver or coin, if necessary.

4. Remove the filter strainer (P2) from the filter receptacle (P3), 

and clean it under running water. Use a soft brush, if 
necessary.

5. If damaged: Replace the filter strainer (P2).
6. Re-assemble the filter cartridge (P) in reverse order.
7. Check the tightness of the filter cartridge (P).

Fig. 23

Filter

14.3 Cleaning the flow armature

14.3.1 Removing the measuring cell

1. Close the water inlet, and switch off the measuring cell.

– Close the shut-off spindles of the sample water inlet (V1) 

and outlet (V2).

– Close and/or switch off the device controlled (for instance a 

gas dosing unit or dosing pump).

– Wait until the display of the measuring amplifier is at zero.
– Switch off the electricity supply to the measuring amplifier.

2. Empty the measuring cell.

– Open the bleeding spindle (L) and deaeration spindle (J), 

and drain the water.

– Unscrew the connections to the sample water inlet (U1) and 

outlet (U2).

3. Remove the electrodes and sensors.

– If a water sensor is fitted: Unscrew the screw (O1) on the 

water sensor (O), and remove the sensor.

– Unscrew the plug for the measuring electrode (E), and 

remove the electrode.

– Remove the reference electrode (B1). Fit thread protective 

caps, and clean the electrode.

– Remove the pH (A1) and redox electrodes (A2), if 

necessary. Fit thread protective caps, and clean the 
electrodes.

– Fit protective caps on the electrodes. The protective caps 

must be filled with 3-molar potassium chloride solution.

– Store electrodes with protective caps in a dry place.

4. Unscrew the measuring electrode.

– Unscrew the screwed part (H) from the flow armature (F) 

with the slide (G1), and remove them carefully.

– Keep O-rings (W1).
– Remove the cleaning wing (X) and inlet chamber (W).

5. Remove the flow armature.

– Slacken the screws of the flow armature (do not unscrew 

completely), press the flow armature upwards, and remove 
it from the mounting plate. 

Fig. 24

Exploded view of screwed part

Caution

Prior to maintenance work: Switch off the system 
as described in section 13.3 Switching off!

Note

After cleaning, check the sample water flow rate, 
and adjust it, if necessary. 

TM

03

 5

880

 1

1

0

6

P1

P2

P3

Caution

The measuring cell must only be removed by 
authorised personnel.

Caution

Observe the installation and operating 
instructions of the electrodes!

Note

Carefully remove the screwed part (H)! The 
cleaning wing (X) and inlet chamber (W) are loose 
on the measuring electrode (D) and may fall out! 
Keep O-rings!

T

M

03

 58

9

2

 1

1

06

G1

D

X

W

H

W1

W1

Содержание AquaCel AQC-D3

Страница 1: ...GRUNDFOS INSTRUCTIONS AQC D3 Measuring cell Installation and operating instructions...

Страница 2: ...006 A1 2009 A2 2009 EN 61000 3 3 2008 Convient uniquement aux produits avec tension de service 50 VAC ou 75 VDC IT Dichiarazione di Conformit CE Noi Grundfos dichiariamo sotto la nostra esclusiva resp...

Страница 3: ...leriyle uyumlu oldu unun yaln zca bizim sorumlulu umuz alt nda oldu unu beyan ederiz D k Voltaj Y netmeli i 2006 95 EC Kullan lan standart EN 61010 1 2001 ikinci bask EMC Diretifi 2004 108 EC Kullan l...

Страница 4: ...ing amplifier 15 12 5 Electrical connections 16 12 6 Checks prior to start up 17 12 7 Switching on 17 12 8 Calibrating the parameters Cl2 ClO2 O3 17 13 Operation 19 13 1 Function 19 13 2 Operation 19...

Страница 5: ...English GB 5 3 Installation sketch...

Страница 6: ...ll with sensor interface Fig 3 AQC D3 potentiostatic measuring cell with DIP measuring amplifier TM03 5894 2611 TM03 5895 2611 TM03 5896 2611 1 Conex DIA 1 Conex DIA 2 Conex DIA 2Q Conex DIS D measuri...

Страница 7: ...c diaphragm with buffer solution PCX pH ceramic diaphragm without buffer solution PTB pH PTFE diaphragm with buffer solution PTX pH PTFE diaphragm without buffer solution PKB pH KCl filling with buffe...

Страница 8: ...pressure loading valve Electrodes for disinfection AU Gold PT Platinum X No disinfection measurement Other electrodes PCB pH ceramic diaphragm with buffer solution PCX pH ceramic diaphragm without bu...

Страница 9: ...armature G1 Slide I Bottom of measuring chamber K Sample water extraction device with extraction spindle L Bleeding spindle M Measuring tube N Float body O Water sensor optional O1 Screw P Filter cart...

Страница 10: ...roller installed Preassembled system with Conex DIA 1 measuring amplifier and controller Measurement free chlorine Cl2 as an option with pH compensation chlorine dioxide ClO2 ozone O3 pH temperature C...

Страница 11: ...electrode 10 2 1 AquaCell AQC D3 potentiostatic measuring cell Cables cable for reference electrode pH or redox single rod electrode 3 m with free cable end cable for measuring electrode 3 m with free...

Страница 12: ...drilling diagram Fig 6 Drilling diagram for AQC D3 measuring cell Fig 7 Drilling diagram for preassembled systems Example with Conex DIA wall mounted unit TM03 5898 2411 300 260 360 395 121 TM03 5899...

Страница 13: ...e mounting plate from factory Individual device for AQC D3 measuring cells 1 Drill four dowel holes with a diameter of 10 mm and min 60 mm deep See section 10 4 Dimensional sketches drilling diagram 2...

Страница 14: ...ol output remains active The advantage of this is that in the event of frequent peaks in the sample water flow rate the alarm is not continuously triggered and the control output remains active If the...

Страница 15: ...odes or electrodes A1 A2 B1 1 Cut the electrode cables to the desired length approximately 80 mm for the connections 2 Remove 80 mm of the outer insulation 3 Disentangle the braided screen until you r...

Страница 16: ...mounting in a control panel Fig 19 Conex DIA 1 DIA 2 DIA 2Q for wall mounting TM03 5870 1106 Pin Component Pin 1 brown Measuring electrode M Pin 2 white Counter electrode G C Pin 3 blue Pt100 tempera...

Страница 17: ...IS D If not done yet Define the basic settings for the measuring amplifier Calibrating the pH value DIP Select the right hand display for pH with the Select button Conex DIA and DIP Use the CAL button...

Страница 18: ...displays with DIP The data of the last 10 calibrations can be read 12 8 3 Calibrating the parameters chlorine chlorine dioxide or ozone with Conex DIS D 1 Press the CAL selector and select the calibr...

Страница 19: ...ntinuously cleaned of deposits by a cleaning wing which has a gentle yet effective action This ensures uniform sensitivity for the measuring cell over a long period The cleaning wing is driven hydraul...

Страница 20: ...d for instance a gas dosing unit or dosing pump if necessary 13 4 2 After long term stop inspection If the measuring cell has been switched off for a long period of time or has been emptied for instan...

Страница 21: ...these cannot be measured if this procedure is used f Incorrect photometric reference measurement during calibration Check the reference measurement and take another measurement Repeat the calibration...

Страница 22: ...lectricity supply to the measuring amplifier 2 Empty the measuring cell Open the bleeding spindle L and deaeration spindle J and drain the water Unscrew the connections to the sample water inlet U1 an...

Страница 23: ...re deposits 2 Clean the filter strainer P2 under running water Use a soft brush if necessary 3 Carefully clean all plastic parts with soap suds 4 Use hydrochloric acid 5 10 for deposits 5 Carefully ri...

Страница 24: ...321 141 Cable for measuring electrode 3 m 96609172 45 10212 1 Water sensor with cable 1 m 96609173 45 10212 Water sensor with cable 3 m 96609174 314 605 Reference electrode 96609175 553 1561 Measuring...

Страница 25: ...25...

Страница 26: ...26...

Страница 27: ...2 Telefax 39 02 95309290 95838461 Japan GRUNDFOS Pumps K K Gotanda Metalion Bldg 5F 5 21 15 Higashi gotanda Shiagawa ku Tokyo 141 0022 Japan Phone 81 35 448 1391 Telefax 81 35 448 9619 Korea GRUNDFOS...

Страница 28: ...es it possible Innovation is the essence The name Grundfos the Grundfos logo and the payoff Be Think Innovate are registrated trademarks owned by Grundfos Management A S or Grundfos A S Denmark All ri...

Отзывы: