NEEDLES DO NOT SEEM TO PENETRATE THE SKIN ENOUGH 
 

This can happen in either of these cases: 

 

(1) The number of needles in the cartridge has been incorrectly set to "1" instead of 9 
or 12 or vice versa. Verify that that question was answered correctly, to the machine. 

That question much be re-answered whenever the cartridge type is changed, so you 

must turn the machine off so it will ask you that question again. 

 

(2) Due to its low duty cycle, our machine is MUCH less painful than others and if you 
used numbing cream then you'll definitely feel nothing on most depth settings, likely 

not at all. Also due to the low duty cycle (very short time inside the skin of the needles), 

the eye is unable to see the 1-needle come out. 

 

(3) A delay in reaching the set depth is by design, since the needling depth is achieved 

by a magnet, hammering on the cartridge and the force of the magnet is carefully 
calibrated in a sensor feedback loop. 

That can only be done safely and accurately when the starting force starts at zero and 
slowly ramps up force and depth over the first few seconds - needle punches, in fact. At 

the FAST speed, this takes only about 0.8 seconds, or 20 punches. At MEDIUM speed, 

this takes about 3 seconds and at SLOW speed, up to 20 seconds. 

So when you feel nothing on SLOW or MEDIUM, that's normal and by design, the first 

few seconds, especially on the "SLOW" setting. The correct needling depth force is still 

being calibrated to your skin, for safety. 

 

(4) Also see “NEEDLES DO NOT COME OUT AT ALL OR SCRATCH” on the previous page. 

 

Next page: HELP US DIAGNOSE THE PROBLEM, IF YOU CAN’T FIND ITS CAUSE. 

 
 

Содержание Derminator 2

Страница 1: ...sed on the older machine and vice versa owndoc shop is the only place that sells correctly working cartridges for this machine Derminator 2 info page with videos and link to store http derminator com The latest version of this user manual http derminator com Derminator2 instructions pdf How to properly attach the rubber feet to the machine http derminator com Mounting the rubber feet pdf ...

Страница 2: ...RTRIDGE MOUNTING 8 HAND UNIT CLEANING 8 CARTRIDGE DISINFECTION 9 DERMINATOR NEEDLING TECHNIQUE 10 THE DISPLAY EXPLAINED 11 TIMER FEATURE FOR OPTIMAL TREATMENT 14 MANUALLY CALCULATING TREATMENT TIME 16 COMPLIANCE STATEMENTS 17 FCC 17 CANADA 17 EU 17 WARRANTY POLICY 18 CARE MAINTENANCE 19 ERROR MESSAGES 20 MAGNET BAD STUCK OR BAD CARTRIDGE 20 MAGNET REMOVED OR ATTRACTED OUT 21 MAGNET FLEW OUT FIND I...

Страница 3: ...TCH 23 NEEDLES DO NOT SEEM TO PENETRATE THE SKIN ENOUGH 24 TECHNICIAN S MENU 26 Firmware version 26 Magnet monitor 26 Swap warning 26 Show Hall value 26 MagnetMissingCnt 26 Dlt HallTreshold 26 StartPulseLength 26 HallMarginFFNPls 26 MinPulseLength 27 FastFreq 27 FastDelay 27 LCDDelay 27 Recalibrate 27 Factory Reset 27 ...

Страница 4: ...medicinally active chemicals that interfere with skin healing Skin products used during or shortly after dermaneedling should be critically examined for potentially troublesome ingredients Needle cartridges should be inspected before use on the presence of bent needle tips The same needle cartridge should never be used on different people and cartridges should always be sterilized directly after u...

Страница 5: ...on NO MICROTEARING To avoid microtearing a very powerful motor is a necessity in order to make the needles move in out extremely quickly and making the machine run on a rechargeable battery would therefore be impractical and very expensive Competing Chinese made devices are based on permanent makeup hardware and are much too feeble to be used for safe dermaneedling They do not reach the advertised...

Страница 6: ...he Korean mCure and they all caused extreme tearing We hope others will repeat our experiment with our machine and other machines and post their results online Microtearing rips the edges of pores causing permanent pore enlargement Only a very powerful motor can move the needles in out in a few milliseconds That is why the Derminator 2 is much louder than Chinese pen devices and has a much larger ...

Страница 7: ... firmly Especially initially use one or more size shorter needle length and one week longer interval than used with manual needling when using the Derminator 2 Pushing less hard will make the needles penetrate less Do not use the timer function initially since it uses an algorithm for a very thorough medically optimal for regeneration treatment Only use the highest repetition rate and or longer ne...

Страница 8: ... maximum resistance CARTRIDGE LIFETIME The question how long do the needle cartridges last is unanswerable since this depends on how many times the cartridge has been plunged into the skin which depends on needling frequency skin area treatment frequency and the thickness of the particular skin All we can say is that generally cartridges can be reused one or more times and specifically that they h...

Страница 9: ...soapy solution has enveloped the needles Keep moving and rinsing until the detergent has been flushed away also from inside the cartridge Then pour some Chloramine T solution or Isopropanol 70 to 100 into a cup and add the cartridges to it Make circling motions with the cup so that the solution swirls around the cartridges Stick your fingers into the cup and operate the cartridges plungers plunger...

Страница 10: ... lines otherwise needling density will not be uniform but stripey As long as you keep moving in circles it s not so important how large those circles are or how fast you re circling or how quickly you re navigating around the skin The only really important thing is to cover all of the to be treated skin more or less equally But you could and perhaps should put increased focus on skin that needs ex...

Страница 11: ... can specify a skin area size and the machine calculates how much time should that area be needled with the speed setting s used When you change the needling speed the remaining time changes as well The most common mode of operation is the Fast setting The left gray button is used to select operating mode ...

Страница 12: ...2 times more pricks than it should when the user has told the machine that a single needle cartridge has been mounted but in fact a 12 needle cartridge is mounted This error can result in permanent disfigurement This setting is also VERY important to 2 Inform the sensor of the needle length of the cartridge used otherwise the machine will needle 1 mm too deep when the user has told the machine tha...

Страница 13: ...has reached zero All the user has to do is move the device with circular motions equally over the entire patch at least once during that time period 6 Needling depth indicator The bottom triangle indicates the set depth in increments of 0 25 mm The deepest needling setting 2 5 mm shares its position with depth setting 2 25 mm The top triangle shows the actual needling depth or more accurately phra...

Страница 14: ... frequency meanwhile The Derminator 2 knows how many pricks need to be made on the skin and when the skin patch has that number of pricks it stops That s all there is to it And when you want to be conservative you just stop sooner This automatic mode is done via the Timer function Timer mode is entered by pressing the right blue Depth button marked OK when the display shows Timer The device then a...

Страница 15: ...hat and will turn itself off as soon as the optimal number of pricks have been made The device will stop needling and show ENDED When the timer has been set by selecting OK to the final value you can start needling The machine will stop automatically when the needle density has been achieved Of course you ll have to actually needle the skin all that time and not take relatively long breaks otherwi...

Страница 16: ...m using fast speed Treatment in seconds Area in cm x 10 the nr of needles in the cartridge How exactly you move the device over the skin during that time and how many times the skin is covered by the device is up to you but it is of course important that the entire area is more or less equally covered resulting in a uniform redness Circular motions are the best needling technique with this device ...

Страница 17: ...cian for help IMPORTANT Any changes or modification not expressly approved by the party responsible for compliance could void the user s authority to operate the device Where shielded interface cables have been provided with the product or specified additional components or accessories elsewhere defined to be used with the installation of the product they must be used in order to ensure compliance...

Страница 18: ...o diagnose an issue An exception to this warranty policy is hand units with internal liquid damage or cracked hand unit shroud due to using chemical disinfectants on it When we can not reproduce a failure or when user error is to blame such as a missing magnet we will neither refund nor repair replace at our cost and that includes any previously refunded return postage You can reach our support de...

Страница 19: ...The outside of the hand unit should never come into contact with anything else but a cloth slightly wet with lukewarm water and a tiny bit of ordinary dishwashing hand soap Alcohol based disinfecting products or any other products that contain organic solvents or dissolved chemicals will crack the hand unit shrouds Never spray any liquids into the hand unit and never spray any liquids onto it Neit...

Страница 20: ...agnet has a problem Try again with a new cartridge and select the CORRECT nr of needles used Case 3 The magnet is glued stuck by blood or gliding serum Clean the hand unit and magnet as shown in https youtu be aTLGUPS czk and try again Case 4 The magnet is present but is the machine being used on hair If you did feel the magnet pushing then you can see these error messages in case the machine is b...

Страница 21: ...ing the new one If you mistakenly have TWO magnets in the hand unit then that is invisible to the naked eye but cartridges will unable to be secured onto the hand unit s tip If a replacement magnet does not solve the issue we ll replace your machine MAGNET REMOVED OR ATTRACTED OUT Cause The reciprocating magnet in the hand unit was there when we shipped the machine to you but for some reason the m...

Страница 22: ... DISPLAY WORKS Blood or gliding serum may have entered the hand unit and dried up glueing the reciprocating magnet shut into its tube Watch this video https youtu be aTLGUPS czk on how to prevent this by cleaning the hand unit and magnet after a messy needling session To get a glued magnet out Slam the hand unit with the tip down hard onto a wooden surface This will dislodge the magnet but it may ...

Страница 23: ...ify that the Silicone spring shroud looks OK The top cartridge has a problem The end of the silicone spring is crumpled up and missing a few mm of straightness This is caused when you push the plunger to make the needles come out and you push a little too far hard This is unfixable the cartridge has become bad and has to be discarded ...

Страница 24: ...nable to see the 1 needle come out 3 A delay in reaching the set depth is by design since the needling depth is achieved by a magnet hammering on the cartridge and the force of the magnet is carefully calibrated in a sensor feedback loop That can only be done safely and accurately when the starting force starts at zero and slowly ramps up force and depth over the first few seconds needle punches i...

Страница 25: ...lease needle a stretched taut piece of paper from the edge of a table with one hand while needling with the other hand with the single needle cartridge on 1 mm and hold it against the light and report whether you see holes Remember this will only work when first turning the machine off and telling it a 1 needle cartridge has been mounted E Another test you should definitely do but it will destroy ...

Страница 26: ...gnet is present flew out is stuck or otherwise has a problem You can set it to OFF when you want to avoid annoying messages when needling the scalp SWAP WARNING When turned OFF the machine will not warn about the importance of setting the correct number of cartridge needles A convenient speedup for clinicians SHOW HALL VALUE Shows the magnetic sensor s measured value Usually when it is closest to ...

Страница 27: ...asked to by our support LCDDELAY Never change this value unless asked to by our support RECALIBRATE If we re ever start selling different magnets or different cartridges the user can recalibrate the machine by mounting a sacrificial 9 or 12 needle cartridge and following the machine s instructions FACTORY RESET Restores the machine to its original settings EXCEPT any recalibrations that were done ...

Отзывы: