BASIC OPERATION - F4
REV E
–03/01/2017 13:38:00
Page 5 of 6
W:\Electrical\INSTRUCT\INSTRUMENT\WATLOW\F4\OBS\BASIC OPERATION - F4 REV E.doc
I.
HOW TO AUTOTUNE
***NOT REQUIRED FOR NORMAL OPERATION
1.
Once oven is powered on, load parts into oven.
2.
Press or until cursor is next to
SP1
on lower display. Press to
select
SP1.
3.
Press or to change current set point to desired set point. Press
to accept change
4.
Press or until cursor is next to
Go to Operations
. Press to
select
Go to Operations
.
5.
Press or until cursor is next to
Autotune PID
. Press to select
Autotune PID
.
6.
Press to select
Channel 1 Autotune
.
7.
Press or until cursor is next to
PID Set 1
. Press to select
PID
Set 1
.
8.
Press
twice to return to
Main Page
. Controller will now perform
Autotune. Once complete, the controller will fill in the values in the PID Set 1.
J.
HOW TO FORCE ON EVENT OUTPUTS
***NOT REQUIRED FOR NORMAL OPERATION
1.
Press
or
until cursor is next to
DigitalOut
on lower display.
2.
Press
to enter
DigitalOut.
3.
Press
or
until the cursor is next to the desired
Event Output
.
Press
enter desired
Event Output
.
4.
Press
or
until the cursor is next to
On
.
Press to accept.
5.
Once completed, press once to return to
Main Page
.