Page 9
MantaRay Portable Area-Velocity Flow Meter
24 HR LOG
Press
from the MAIN display to view a formatted flow report from instruments
with a built-in data logger. Press
to pan through
Level
,
Velocity
and
Flow
summaries. Press
to scroll down one day or repeatedly to scroll to a
specific date. Up to 365 days can be stored. Newest date will overwrite the oldest.
Press
to return to the main display.
PASSWORD
The
Password
(a number from 0000 to 9999) prevents unauthorized access
to the
Calibration
menu.
From the Main display press
to get to
Password
. Factory default
password is 0000 and if it has not been changed press
to proceed to the
Menu
Selections
screen.
If a password is required, press
to place the cursor under the first digit and
or
to set the number, then
to the second digit, etc. Press
or
to proceed to
the
Menu Selections
screen.
A new password can be stored by going to
Special Functions/New
Password.
--24 hr log------Flow
Date
Feb. 12/2010
Total
50138 USG
Average
34.82 USG/m
Maximum
52.20 USG/m
Max Time
11:08:00
Minimum
0.000 USG/m
Min Time
9:15:00
}
--Password----------
Password
0000