Page 20
DFS 5.1 Doppler Flow Switch
APPENDIX B
FLOW VELOCITY CALCULATION
Flow velocity can be calculated from the following data.
KNOWN FLOW
DIVIDED BY
VELOCITY
U.S. gallons/sec.
0.0408 x (pipe I.D.")²
= ft/sec
U.S. gallons/min.
2.448 x (pipe I.D.")²
= ft/sec
Imp. gallons/sec
0.0340 x (pipe I.D.")²
= ft/sec
Imp. gallons/min
2.040 x (pipe I.D.")²
= ft/sec
Litres/sec
0.1545 x (pipe I.D.")²
= ft/sec
Litres/min
9.270 x (pipe I.D.")²
= ft/sec
Cubic metres/sec
0.0001545 x (pipe I.D.")²
= ft/sec
Cubic metres/min
0.00927 x (pipe I.D.")²
= ft/sec
EXAMPLE: 3" ID pipe
Known Flow of 160 US gallons/min: 160 = 7.26 ft/sec
2.448 x (3)²