iCFilm

User Guide

Содержание iCFilm

Страница 1: ...iCFilm iCFilm User Guide...

Страница 2: ...percentage 10 8 6 Measure soft dot 11 8 7 Measure negative soft dot 12 9 Calibration 13 9 1 Calibration check 13 9 2 Calibration procedure 14 10 Adapt iCFilm to other measurement devices 15 10 1 Condi...

Страница 3: ...ect sunlight The iCFilm should be protected against chemical substances corrosive vapors strong vibrations and mechanical influences The iCFilm should always be transported in the original packaging T...

Страница 4: ...UserGuide E N G L I S H 3 GretagMacbeth iCFilm User Guide E N 2 Package contents 1 iCFilm 2 Calibration wedge 3 Serial cable 4 Quick start guide 5 Device certificate 6 Registration card 7 Device case...

Страница 5: ...UserGuide 3 Packaging and transportation The iCFilm should always be transported in the original packaging UserGuide E N G L I S H 4 GretagMacbeth iCFilm User Guide E N...

Страница 6: ...rGuide E N G L I S H 5 GretagMacbeth iCFilm User Guide E N 1 2 3 7 4 6 8 5 1 Display 2 Serial interface 3 Button A 4 Button B 5 Measurement aperture 6 Locking button 7 Batteries 8 Storage compartment...

Страница 7: ...g any button Lock measuring head Fold up the iCFilm and push in the locking button The display goes out after approx 30 seconds 6 Buttons UserGuide E N G L I S H 6 GretagMacbeth iCFilm User Guide E N...

Страница 8: ...s UserGuide E N G L I S H 7 GretagMacbeth iCFilm User Guide E N DENS Density DOT Screen percentages NEG DOT Negative screen percentages SOFT DOT Soft screen percentages NEG SOFT DOT Negative soft scre...

Страница 9: ...ged in the compartment on the bottom of the unit Before you put the iCFilm in operation you should select the desired aperture Position the aperture in the let in snug fit intended for it on the basep...

Страница 10: ...n A and simultaneously push the iCFilm down Do not release the measurement head or button A before 0 00 0 0 or 100 0 is displayed Note The last zero reference is stored in the memory of the iCFilm eve...

Страница 11: ...m down until 100 0 is displayed Press shortly button A Place the required dot area patch on the measuring aperture Press the iCFilm down until the digital value is displayed Any amount of dot percenta...

Страница 12: ...ue 50 dot area i e UGRA Plate Control Wedge RIT Microline Resolution Target etc Zero the device see section 8 2 Select SOFT DOT function by using button B as often as necessary Select the adjustment m...

Страница 13: ...d until NEG SOFT DOT ADJ is displayed Press shortly button A Place the required negative dot area patch on the measuring aperture Press the iCFilm down until the digital value e g 46 0 is displayed Us...

Страница 14: ...dge may lead to a faulty calibration and therefore incorrect measuring results 9 1 Calibration check The calibration check of the iCFilm should be performed periodically using the GretagMacbeth calibr...

Страница 15: ...d keep them pressed until the current slope factor is displayed original 1000 Press shortly button A Place the density patch of the calibration wedge onto the measurement aperture Press down the iCFil...

Страница 16: ...performed with reference either to a required density value or to a given slope factor if known The proceeding is the same as mentioned in section 9 for the calibration 10 1 Conditions for adapting th...

Страница 17: ...button B to increase the displayed value To cancel ADJ DENS mode press buttons A and B simultaneously until DENS appears The two densitometers are now compatible and the new slope factor can be read o...

Страница 18: ...he polarity into account The polarity and the mounting position are illustrated in the battery compartment Put the battery cover back on again Dispose of the old batteries in accordance with regulatio...

Страница 19: ...mends annual re certification of the iCFilm Transmission densitometer Contact your GretagMacbeth dealer or your nearest GretagMacbeth Service Centre for more information on the recertification process...

Страница 20: ...d Neg Soft Dot Max decrease at 50 10 Linearity Repeatability 0 01 D Precision 0 5 Calibration Slope for density measurements Measurement geometry Specular diffuse Apertures 1 mm 2 mm 3 mm Length of me...

Страница 21: ...to the serial interface socket remove the black cover on the rear of the instrument head The serial interface is configured as follows in the factory Format Standard RS 232 Data rate 1 200 baud Data f...

Страница 22: ...cluding all applicable amendments 73 23 EEC Electrical equipment for use within specified voltage limits 89 336 EEC Electromagnetic compatibility and the standards and or technical specifications refe...