Z31.0.XY.6C-09
Operating manual GTF 102-Ex... und GTF 101-Ex...
page 4
_______________________________________________________________________________________________________________________
Zone 1, 2 or 21, 22
A GTF 101-Ex or GTF 102-Ex complying with the protection class EEx em II T6 or EEx ia IIC T6
can be used, if the temperature has to be detected in EX-zone 1, 2 or 21, 22. In case the
controlling- or displaying device is also in the zone 1, 2 or 21, 22 only ATEX certified devices
complying with the demands of the certain zone are allowed to be used.
In case the controlling- or displaying device is beyond the explosion hazardous area, this device
does not need an Ex-certification if a temperature-probe GTF 101-Ex or GTF 102-Ex complying
with the protection class increased safety (EEx em II T6) is used and the safety instructions given
in chapter 2 are fulfilled.
Table 1 : Overview of application and demands to a GTF 102-Ex... or a GTF 101-Ex and the display-/ control-
device in different Ex-zones in detail.
GTF 101-Ex and GTF 102-Ex type-code are shown in chapter “ Specification”
Ex-zone: sensor
0 , 20
0/1 , 20/21
1, 2, 21, 22
1, 2, 21, 22
Ex-zone: device
0, 20
1, 2 ,21, 22
1, 2, 21, 22
no Ex-environment
Applicable
GTF 101-Ex
GTF 102-Ex
Types
nonlicenced
(see GTF 103-Ex)
GTF 102-Ex...
GTF 101-Ex...
GTF 102-Ex...
GTF 101-Ex...
GTF 102-Ex...
ATEX-marking
of the sensor
II 1GD 80°C
EEx ia IIC
Ta= -20°C...+40°C
II 1/2 GD 80°C
EEx ia IIC T6
II 2 GD 80°C
EEx ia IIC T6 or
EEx em II T6
II 2 GD 80°C EEx em II T6
ATEX standards
to the device(s)
II 1 G resp.
II 1 D
II 1G; II (1)2G EEx ia IIC,
II 2G EEx ia IIC bzw.
II 1D; II (1)2D EEx ia IIC,
II 2D EEx ia IIC
II 2 G resp.
II 2 D
without ATEX-admission