GREISINGER GOX 91 Скачать руководство пользователя страница 4

After re-positioning, rinse membrane with water. We highly recommend to wait at least 30 to 45 minutes
before starting to re-calibrate the unit. In order to make absolutely sure that measurements will be correct, allow
electrode to stabilise for 12 hours (during the night) before re-calibrating as Teflon material, which is easily
overstretched during positioning, will without fault return to its original state after some time.
Once this state has been reached again, this can be seen at the display of the unit as immediately after replacing
the membrane a relatively high value is displayed; due to the tension in the material being relieved the value on
the display will fall accordingly.

Please note that our experience has shown that in most cases inaccurate measurements are the results of the
membrane being incorrectly replaced and/or the unit being incorrectly calibrated. Please make sure to emplay
utmost care when replacing the membrane or calibrating the unit.
In the course of the time and due to temperature differences during measurements the membrane will loose its
elasticity and, if electrode has not been stored in water, will have to be replaced after 3 months regardless of its
being defective or not.

Содержание GOX 91

Страница 1: ...table top or suspended operation Weight Approx 450 g incl battery and electrode Points to be observed prior and during measuring Prior to measuring make sure that the electrode has assumed water temperature To test please proceed as follows switch device to temperature indication and move electrode around in water until stable temperature is displayed The electrode is now ready for oxygen measurem...

Страница 2: ... 6 12 5 16 10 4 26 8 6 36 7 4 7 12 2 17 10 2 27 8 5 8 12 0 18 10 0 28 8 3 9 11 8 19 9 9 29 8 2 Customer may of course choose one of the two methods of calibration at his own discretion simple calibration in air or conventional wet calibration in saturated water Water calibration Although cold water usually absorbs more oxygen it is much more difficult to determine its actual degree of saturation w...

Страница 3: ...ce membrane till electrolyte is showing 9 Turn electrode round holding it vertically pointing upwards 10 Wipe off silver cathode with a paper tissue till clean but do not polish The scratches on the silver surface have been deliberately made as scuffing assists the formation of an electrolyte film 11 Put one drop of electrolyte on the silver plate put new membrane on top and re position first red ...

Страница 4: ... as immediately after replacing the membrane a relatively high value is displayed due to the tension in the material being relieved the value on the display will fall accordingly Please note that our experience has shown that in most cases inaccurate measurements are the results of the membrane being incorrectly replaced and or the unit being incorrectly calibrated Please make sure to emplay utmos...

Отзывы: