Operating Manual GMH 3430 

page 5 of 10

 

 

Selection of Measuring Unit 

The device allows a choice between the various measuring units, i.e. conductivity, resistivity of fluid, filtrate dry resi-
due ('Total Disolved Solids') and salinity. To do so press 

-key for a short time. The measuring value of the current 

measuring unit will be shown in the upper line of the display, the corresponding device will be indicated by an arrow 
at the top corner of the display. Simultaneously, the temperature is shown in the lower line of the display with all 
measuring units.  
 

Change Between Measuring Range Resolutions 

For measuring conductivity you have a choice between four different measuring range resolutions: 

 

0.0  ...  200.0 

S/cm 

 

0  ... 

2000 

S/cm 

 

0.00  ...  20.00  mS/cm 

 

0.0  ...  200.0  mS/cm 

 
If Auto-Range is switched on the GMH3430 will automatically choose the optimum resolution. If Auto-Range is 
switched off you can change to the next measuring range resolution by pressing the 

-key for a short time. 

In case you have already been working in the highest resolution range, pressing of the key will bring you down to the 
lowest measuring range resolution. The corresponding measuring unit is shown by an arrow in the top right-hand 
corner of the display. 

 

To activate/deactivate Auto-Range press 

-key press for 2 seconds.  

The Auto-Range function set will be shown in the display as long as the 

 -key is being 

pressed. 

Salinity measuring: 

The salinity (salt content) of 

seawater can be determined in the measuring mode „SAL“  

(basis: International Oceanographic Tables; IOT) 

The salinity of standard-

seawater is 35 ‰ ( 35g salt per 1kg seawater). 

The values are displayed in ‰ (g/kg). 
 

Display at reboot 

After the device is switched on it will display some information about the current configuration.  
There are the following displays:  

 

segment test 

 

 

set scale correction of measuring cell  

 

 

set compensation coefficient  

(only displayed if t.Cor = lin) 

 

set temperature offset  

(not displayed if OFFS = off) 

 

setting for Autorange 

 

Then 0 is displayed until the first valid measuring value is available.   

 

 

   S

Logg

%/K

AL

mS

nLF

SAL

mg/l

1/cm

Lin

kOhm

Содержание GMH 3430

Страница 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de Measuring device for conductivity temperature as of version V2 1 Operating Manual GMH 343...

Страница 2: ...SURING CELL 8 13 ERROR AND SYSTEM MESSAGES 8 14 THE SERIAL INTERFACE 9 14 1 THE FOLLOWING INTERFACE FUNCTIONS WILL BE SUPPORTED 9 15 SPECIFICATION 10 16 DISPOSAL INSTRUCTION 10 1 Designated Use The de...

Страница 3: ...entical to the mains voltage 4 Safety Requirements This device has been designed and tested in accordance with the safety regulations for electronic devices However its trouble free operation and reli...

Страница 4: ...Configuration to enter values or change settings Range for conductivity measuring only press for 2 sec change between automatic and manual measuring range selection press shortly at manual measuring...

Страница 5: ...n by pressing the key for a short time In case you have already been working in the highest resolution range pressing of the key will bring you down to the lowest measuring range resolution The corres...

Страница 6: ...C t Lin Setting of Temperature Coefficient only when t Cor Lin 0 300 3 000 Temperature compensation coefficient in K t rEF Selection of Reference Temperature only when t Cor nLF or Lin 20 C 68 F refe...

Страница 7: ...hannel 3 23 Using the interface converter GRS3105 it is possible to connect several devices to a single interface As a precondi tion the base addresses of all devices must not be identical In case sev...

Страница 8: ...tem Messages Error or system messages Description Reason Remedy Low battery voltage device will only continue operation for a short time replace battery Low battery voltage replace battery If mains op...

Страница 9: ...a Programming examples Visual Basic 4 0 Delphi 1 0 Testpoint If you want to use the interface functions make sure to switch off the Auto Range function 14 1 The following interface functions will be s...

Страница 10: ...10 5 12V direct voltage supply suitable power supply GNG10 3000 Power consumption approx 3 5 mA Low battery warning and bAt Display 2 four digit LCD displays 12 4mm and 7mm high for meas values min an...

Отзывы: