3 Operation Of Logger
The device supports two different logger functions:
„Func-Stor“:
each time when „store“ (key 6) is pressed a measurement will be recorded.
„Func-CYCL“:
measurements will automatically be recorded each interval, which was set in the logger
menu ‚CYCL‘ until the logger will be stopped or the logger memory is full. The recording is
started by pressing „Store“ 2 seconds.
The logger records 3 measurement results each time:
current or mean value (depending on logger setting, see below), min peak and max peak.
Min and max peak
are the minimum resp. the maximum of the measured values since the last recording.
Using them allows f.e. analysis of fluctuating pressures.
For the evaluation of the data the software GSOFT3050 has to be used. The software also allows easy
configuration and starting of the logger.
When the logger is activated (Func Stor or Func CYCL) the hold function is no more available, the key 6 is
solely used for the operation of the logger functions.
3.1 „Func-Stor“: Storing Single Measurements
Each time when „store“ (key 6) is pressed a measurement and its time stamp will be recorded.
The recorded data can be viewed either in the display (when calling the configuration an additional menu
„REAd LoGG“ is displayed, see below) or by means of the interface and a PC with GSOFT3050-software.
Max. number of measurings:
99
A measuring contains:
- current measuring value at the time of recording
- min peak, max peak since the last recording
- time and date of the recording
After each recording „St. XX“ will be displayed for a short time. XX represents the number of the recording.
If logger memory contains recordings already:
When „Store“ is pressed for 2 seconds, the choice for clearing the logger memory will be displayed:
The selection can be made by
5
(key 2) and
6
(key 5). "Quit" (key 6) enters the choice.
If the logger memory is full, the display will show:
Viewing Recorded Measurings
Within the „LoGG Stor“ function the measurings can be viewed directly in the display not only by means of a
computer (like at „Func CYCL“): press 2 seconds „Set“ (key 4): The first menu displayed now is „rEAd LoGG“
(read logger data). After pressing
4
(key 3) the measurement recorded last will be displayed, changing
between the different values referring to the measurement also is done by pressing
4
.
Changing the measurement is done by pressing the keys
5
or
6
.
Clear all
recordings
Clear the last
recording
Clear nothing
(cancel menu)