E40.0.01.6C-04 

Operating Manual EASY

LOG

 80CL 

Page 3 of 14 

3. Safety advice 

This device has been designed, assembled and tested in accordance with the safety regu-
lations for electronic measuring device. However, its trouble-free operation and reliability 
cannot be guaranteed unless the standard safety measures and special safety advices 
regarding the device will be adhered to when using the device. 

1. Trouble-free operation and reliability of the device can only be guaranteed if the 

device is not subjected to any other climatic conditions than those stated under 
”Specification”. 

To protect the battery, the maximum permissible storage and transport 
temperature of the device is 70 °C

 

2. Standard regulations for operation and safety for electrical, light and heavy current 

equipment have to be observed, with particular attention having to be paid to 
national safety regulations. 

3. When connecting the logger to other devices (e.g. PC) the interconnection has to 

be designed most thoroughly as internal connections in third-party devices (e.g. 
connection GND with protective earth) may lead to undesired voltage potentials. 

4. If there is any risk whatsoever involved in running it, the device has to be switched 

off immediately and to be marked accordingly to avoid re-starting. 

  Operator safety may be at risk if: 
 

• there is visible damage done to the device 

 

• the device is not working as specified 

 

• the device has been stored under unsuitable conditions for a longer time. 

  In case of doubt, please return device to manufacturer for repair and maintenance. 

5. Warning:

 

Do not use these product as safety or emergency stop devices, or in any other 
application where failure of the product could result in personal injury or material 
damage. 

Failure to comply with these instructions could result in death or serious 
injury and material damage.

 

4. 

Notes on the logger’s delivery status 

The delivered logger is in a “sleep mode”, that means nothing is displayed and the 
power consumption is at its minimum. When there is a communication with the software 
for the first time, the EASY

LOG

 

leaves the sleep mode and ›Stop‹ is displayed. Then the 

device is ready for operation. 

Note: 

That sleep mode (nothing is displayed) cannot be restored again. 
“Stop“ or  “Halt“ are similar to the sleep mode. Both of  that operation 
modes ensure minimum consumption. 

Note: 

Logger start over keypad: 
before logger start over keypad, logger time must be adjust with 
GSOFT 40K. 

 

 

 

 

Содержание 602773

Страница 1: ...electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 phone 49 9402 9383 0 fax 49 9402 9383 33 eMail info greisinger de ab f from Version V1 8 Operating Manual EASYLOG 80CL Data logger for humidity...

Страница 2: ...logger EASYLOG 80CL is designed for long term monitoring of climate data Not only relative humidity temperature and air pressure can be displayed but also wet bulb tem perature dew point temperature...

Страница 3: ...tentials 4 If there is any risk whatsoever involved in running it the device has to be switched off immediately and to be marked accordingly to avoid re starting Operator safety may be at risk if ther...

Страница 4: ...play element for max min Logg indicates that logger is recording BAT signals weak battery see chapter 9 AL signals an alarm state at minimum one channel SL signals an active sea level correction for t...

Страница 5: ...the display element MAX Note You can immediately switch to the next measuring channel pushing the button again Delete max value Push button for about 5 seconds CLr is displayed and the max values of...

Страница 6: ...tart recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with the arrow keys button or and verify with Note If no key is p...

Страница 7: ...ed if the logger is stopped logger without logger data 7 4 Operating possibilities via software EBxKonfig or EASYBus Configurator The software EBxKonfig V3 9 or EASYBus Configurator V2 0 is needed to...

Страница 8: ...ow AL signals an alarm state at minimum one channel The arrow SL signals an active sea level correction 9 2 Logger status messages At stand alone operation no EASYBus connected logger status messages...

Страница 9: ...er to the manufacturer for repair 9 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured value is below...

Страница 10: ...at range 10 90 RH Air pressure 1 0 mbar typ 2 5 mbar max Display two 4 digit LCD displays about 7 and 5 8 mm high Interface EASYBus Busload equals to 2 EASYBus elements Battery life time depending on...

Страница 11: ...ng transparent screen made of polycarbonate splash proof acc to IP65 excl protection cap Sensor tube about 15 mm made of polyamide Protection cap removable twist plastic head made of polycarbonate IP4...

Страница 12: ...oftware delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceeded Measuring range of the chan nel has been exceeded Operate logger within...

Страница 13: ...tion diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connected to terminal 10...

Страница 14: ...E40 0 01 6C 04 Operating Manual EASYLOG 80CL Page 14 of 14...

Отзывы: