background image

T14.0.0X.6C-02

User’s Manual GLMU 200 MP

page 7 of 8

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

5.8  'dA.Lo': Display at zero output (output scaling)

Enter the display value at which the output should have 4mA (or 0V).
The value range is depending on the selected measuring function (Func) and range (rAnG).

5.9  'dA.Hi': Display at maximum output (output scaling)

Enter the value at which the output should have 20mA (or 1 / 10V).
The value range is depending on the selected measuring function (Func) and range (rAnG).

5.10  'Unit' with Temp-arrow: Temperature unit

Temperature unit: All referring settings and displays are done in this unit.
Choice between °C and °F (ex works: °C)

5.11  'OFFS' with Temp-arrow: Offset of temperature measuring

The offset of the measuring will be shifted by this value, the input is in °C. Calculation: see below.
Max. input range: -5.0...5.0 °C or 'oFF': offset is deactivated (=0.0, ex works)

5.12  'SCAL' with Temp-arrow: Scale correction of temperature measuring

The scale of the measuring is changed by this value. Calculation: see below.
Max. input range: -5.00...5.00 or 'oFF': scale is deactivated (=0.00, ex works)
The adjusting of the temperature by means of offset and scale is intended to be used to compensate
deviations of the measuring. It is recommended to keep the scale correction deactivated (“oFF”). The display
value is given by following formula:

Temperature display = measured value - offset

With a scale correction (just for calibration laboratories, etc) the formula changes:

Temperature display = (meas. value - offset) * ( 1 + scale correction/100)

After pressing key 1 again, the instrument will restart (display 8888).

6 Disposal instructions

The device must not be disposed in the regular domestic waste.
Send the device directly to us (sufficiently stamped), if it should be disposed. We will dispose the device
appropriate and environmentally sound.

Содержание GLMU 200 MP

Страница 1: ...3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10mA 0 10V or refer to type plate Reverse voltage protect...

Страница 2: ...cluding domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third...

Страница 3: ...t or voltage output is set by works and cannot be changed 3 4 Terminal assignment of meas cell socket contact no Standard Cell K 0 55 1 Temperature 2 Temperature GND 3 Supply 1 4 Signal 1 5 Signal 2 6...

Страница 4: ...linity salt content of seawater can be determined in the measuring mode SAL basis International Oceanographic Tables IOT The salinity of standard seawater is 35 35g salt per 1kg seawater The values ar...

Страница 5: ...al Temperature below 5 C not allowed Choose right measuring range for conductivity Err 7 System fault Error in device Disconnect from supply and reconnect If error remains return to manufacturer Err 9...

Страница 6: ...the conductivity or resistivity range Choice the desired measuring range range is depending on the function set above Possible inputs Range rAnG 0 Range rAnG 1 Range rAnG 2 Range rAnG 3 Conductivity...

Страница 7: ...5 0 C or oFF offset is deactivated 0 0 ex works 5 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input...

Отзывы: