Operating Manual for conductivity-transmitter

GLMU 200 MP

1 Specification

Measuring range 1 (one of the following can be configured):

Range (rAnG)=0

Range (rAnG)=1

Range (rAnG)=2

Range (rAnG)=3

Conductivity

0.0 ... 200.0  mS/cm

0.00 ... 20.00 mS/cm

0 ... 2000 µS/cm

0.0 ... 200.0 µS/cm

Resistivity

5.0 ... 100.0 Ohm·cm

50 ... 1000 Ohm·cm

0.50 ... 10.00 kOhm·cm

5.0 ... 100.0 kOhm·cm

TDS

-

-

0 ... 2000 mg/l

0.0 ... 200.0 mg/l

Salinity

0.0 ... 70.0

-

-

-

Meas. range 2

Temperature:  -5.0...140.0 °C (resolution 0.1°C or 0.1°F, NTC10k)

Attention: Consider application range of measuring cell: 0..80°C

Accuracy:

  (at nominal temperature = 25°C)

  Measuring:

conductivity: 

±0,5% of measured value ±0,3 FS

temperature:

±0.1°C of measured value ±1 Digit

  Add. output signal:

±0.2 % FS

Measuring cell:

4 electrode measuring cell with temperature sensor NTC10k,
7 pole connector
Cell constant K: 0.45..0.65, determined and adjusted ex works

Temperature compensation:

  non linear function, linear function (coefficient configurable) or none

Min-/Max-Value Memory:

Min and max measured values are stored

Output signal:

refer to type plate, freely scaleable

Scaling:

By entering display values for 4mA (or 0V) and 20mA (or 1V/10V) output

Connection:

4 - 20 mA  (2-wire) – output signal electrically isolated

 for option AV01, AV10:

0 - 1 (10) V  (3-wire) – output signal electrically isolated

Auxiliary energy:

 (supply voltage)

Uv = 12 - 30 V DC (4-20mA)
Uv = 12 - 30 V DC, max. 10mA (0-1V)

Uv = 18 - 30 V DC, max. 10mA (0-10V)

or refer to type plate

Reverse voltage protection:

50V permanent

Perm. impedance 

(at 4-20mA)

:

  RA(Ohm) < ( (Uv - 12V) / 0.02A )

Example: for Uv = 18V: RA < (18V - 12V) / 0.02A     =>   RA < 300 Ohm

Permissible load 

(at 0-...V)

:

   RL(Ohm) > 3000 Ohm

Adjusting:

 

via keypress. Conductivity by editing cell constant K (CELL)
Temperature by offset and scale.

Display:

  

approx. 10 mm high, 4-digit LCD-display

Ambient conditions for electronics:

Nominal temperature:

25°C

Operating condition:

-25 to 50°C,  0 to 95 %RH (non-condensing)

Storage temperature:

-25 to 70°C

Ambient of standard measuring cell

-5 to 80°C (short time up to 100°C)

Housing:

ABS (IP65)

Dimensions:

82 x 80 x 55 mm (without elbow-type plug and sensor tube)

Mounting:

With holes for wall mounting (in housing - accessible after cover has been removed).

Mounting distance:

50 x 70mm, max. shaft diameter of mounting screws is 4 mm.

Electrical connection:

elbow-type plug conforming to DIN 43650 (IP65),
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm

EMC:

In accordance with EA1 +A2 (appendix A, class B), additional errors: < 1% FS.
When connecting long leads adequate measures against voltage surges have to be taken.
The electrode connections have to be protected sufficiently against ESD pulses, if the
device is used in areas with risk of ESD.

GREISINGER 

electronic GmbH

Содержание GLMU 200 MP

Страница 1: ...3 wire output signal electrically isolated Auxiliary energy supply voltage Uv 12 30 V DC 4 20mA Uv 12 30 V DC max 10mA 0 1V Uv 18 30 V DC max 10mA 0 10V or refer to type plate Reverse voltage protect...

Страница 2: ...cluding domestic safety regulations e g VDE have to be observed 3 If device is to be connected to other devices e g via PC the circuitry has to be designed most carefully Internal connection in third...

Страница 3: ...t or voltage output is set by works and cannot be changed 3 4 Terminal assignment of meas cell socket contact no Standard Cell K 0 55 1 Temperature 2 Temperature GND 3 Supply 1 4 Signal 1 5 Signal 2 6...

Страница 4: ...linity salt content of seawater can be determined in the measuring mode SAL basis International Oceanographic Tables IOT The salinity of standard seawater is 35 35g salt per 1kg seawater The values ar...

Страница 5: ...al Temperature below 5 C not allowed Choose right measuring range for conductivity Err 7 System fault Error in device Disconnect from supply and reconnect If error remains return to manufacturer Err 9...

Страница 6: ...the conductivity or resistivity range Choice the desired measuring range range is depending on the function set above Possible inputs Range rAnG 0 Range rAnG 1 Range rAnG 2 Range rAnG 3 Conductivity...

Страница 7: ...5 0 C or oFF offset is deactivated 0 0 ex works 5 12 SCAL with Temp arrow Scale correction of temperature measuring The scale of the measuring is changed by this value Calculation see below Max input...

Отзывы: