T20.0.0X.6C-07
page 3 of 4
Arrow to Temp indicates
temperature display
Design types, dimension
Design type with 4-20mA output or voltage with not isolated outputs
GHTU-1R-MP
GHTU-1K-MP
GHTU-2K-MP
GHTU-MP-Kabel
short sensor tube aside
long sensor tube aside
long sensor tube bottom
separated sensor tube
At design types with electrically isolated voltage outputs there are plugs at both
sides (right plug for humidity, left plug for temperature)
Display functions
Currently measured values
During normal operation the
relative air humidity
in
[%RH]
is displayed alternating to
the
temperature
in
[°C]
or
[°F]
.
display relative air humidity
display temperature
Min/Max Value Memory
watch Min values (Lo): press ‘down‘(2) shortly once
display changes between ‘Lo‘ and Min values
watch Max values (Hi): press ‘up‘(3) shortly once
display changes between ‘Hi‘ and Max values
restore current values: press ‘down‘(2) or ’up’(3) once again current values are displayed
clear Min-values:
press ‘down’(2) for 2 seconds
Min values are cleared. The display shows shortly ‘CLr‘.
clear Max-values:
press ‘up’(3) for 2 seconds
Max values are cleared. The display shows shortly ‘CLr‘.
After 10 seconds the currently measured values will be displayed again.
Error and system messages
Display
Description
Possible fault cause
Remedy
Err.1
measuring range exceeded
Wrong signal
Temperature above 120°C not allowed.
Err.2
Measuring value below
measuring range
Wrong signal
Temperature below -40°C not allowed.
Err.7
System fault
Error in device
Disconnect from supply and reconnect. If error
remains: return to manufacturer
Err.9
Sensor error
Sensor or cable defective
Check sensor, cable and connections
Er.11
Calculation not possible
Calculation variable missing
or invalid
Check temperature
8.8.8.8
Segment test
The transducer performs a display test for 2 seconds after power up. After
that it will change to the display of the measuring.