background image

T20.0.0X.6C-07

page 1 of 4

Operating Manual

Relative Air Humidity And Temperature Transducer

GHTU ... MP

 as of V2.6

Specification:

Measuring range:

relative air humidity:

 0.0...100.0 % rel. humidity (temperature compensated)

temperature: 

-40.0...120.0 °C or -40.0...248 °F

Rec. range of application:

standard:

20,0...80,0 %RH

option high humidity:

  5,0...95,0 %RH

Accuracy:

  (at nominal temperature = 25°C)

  Display

relative air humidity: 

±1% linearity, 2% hysteresis 

(at recommended range of application

)

temperature:

±0.4% of measured value ±0.2°C

  Add. output signal:

±0.2 % FS

Sensors:

relative air humidity:

cap. polymer sensor

temperature: Pt1000

Temperature compensation:

 automatic

Min-/Max-Value Memory:

Min and max measured values are stored

Output signals:

refer to type plate, both freely scaleable

Scaling:

By entering display values for 4mA (resp. 0V)
and 20mA (resp. 1V/10V) output

Connection:

4 - 20 mA  (2-wire) – output signals are
electrically isolated from each other

 for option AV01, AV10:

0 - 1 (10) V  (3-wire) – output signals are not electrically isolated from each other

 for option AV01G, AV10G:

0 - 1 (10) V  (3- resp. 4-wire) – output signals are electrically isolated from each other

Auxiliary energy:

 

(supply voltage, each output)

Uv = 12 - 30 V DC (4-20mA)

Uv = 12 - 30 V DC, max. 10mA (0-1V)
Uv = 18 - 30 V DC, max. 10mA (0-10V)
or refer to type plate

Reverse voltage protection:

50V permanent

Perm. impedance 

(at 4-20mA)

:

   R

A

(Ohm) < ( (Uv - 12V) / 0.02A )

Example: for Uv = 18V:

R

A

 < (18V - 12V) / 0.02A     =>   R

A

 < 300 Ohm

Permissible load 

(at 0-...V)

:

   R

L

(Ohm) > 3000 Ohm

Adjusting:

 

via keypress by input of offset and scale (humidity and temperature)

Display:

  

approx. 10 mm high, 4-digit LCD-display

Ambient conditions for electronics:

Nominal temperature:

25°C

Operating temperature:

-25 to 50°C

(sensor head and tube: -40 to 100°C – for short time up to 120°C)

Relative humidity:

0 to 95 %RH   (non-condensing)

Storage temperature:

-25 to 70°C

Housing:

ABS (IP65)

Dimensions:

82 x 80 x 55 mm (without elbow-type plug and sensor tube)

Mounting:

With holes for wall mounting (in housing - accessible after cover has been removed).

Mounting distance:

50 x 70mm, max. shaft diameter of mounting screws is 4 mm.

Electrical connection:

elbow-type plug conforming to DIN 43650 (IP65),
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm

EMC:

The device corresponds to the essential protection ratings established in the Regulations of the Council for the
Approximation of Legislation for the member countries regarding electromagnetic compatibility (89/336/EWG). In
accordance with EA1 +A2 (appendix A, class B), additional errors: < 1% FS.
When connecting long leads adequate measures against voltage surges have to be taken.

GREISINGER 

electronic GmbH

 Working area of humidity sensor: