background image

980207 Rev.01 10-2009 

 
 

 
 
 
 
 
 
 

VACUETTE

®

 QUICKSHIELD 

Safety Tube Holder 

 

Instructions for use 

 
 

Intended Use

: The VACUETTE

®

 QUICKSHIELD Safety Tube Holder is intended to be used only with  VACUETTE

®

 Multiple Use Drawing 

Needles and VACUETTE

®

 Blood Collection Tubes as a system in routine  venipuncture procedures. This device is to be used  by properly 

trained healthcare professionals only in accordance with these instructions. 
 

Product  Description

:  The  VACUETTE

®

  QUICKSHIELD  Safety  Tube  Holder  is  a  non-sterile  single-use  plastic  tube  holder.  The 

VACUETTE

®

 QUICKSHIELD Safety Tube Holder is designed with a safety shield, which can be activated to cover the needle immediately 

following venipuncture to aid in the protection against accidental needlestick injury. 
 
 

Precautions/Cautions 
 
Precautions 

Check VACUETTE

®

 QUICKSHIELD Safety Tube Holders before use. Do not use VACUETTE

®

 QUICKSHIELD Safety Tube Holders when 

damaged! Any VACUETTE

®

 QUICKSHIELD Safety Tube Holder that becomes contaminated with blood is considered hazardous and must 

be disposed of immediately. Do  not reshield needle.  After  venipuncture, dispose of used needle with VACUETTE

®

 QUICKSHIELD  Safety 

Holder as one unit in appropriate sharps container in accordance with policies and procedures of your facility. 
Keep hands behind needle at all times during use and disposal. 
 

Caution 

1. 

Handle all biological samples and blood collection “sharps“ (lancets, needles, Luer adapters, and blood collection sets) according to the 
policies and procedures of your facility. 

2.  Obtain  appropriate  medical  attention  in  the  case  of  any  exposure  to  biological  samples (for example, through a puncture injury), since 

they may transmit HIV (AIDS), viral hepatitis, or other infectious disease. 

3. 

Discard all blood collection “sharps” in biohazard containers approved for their disposal. 

4.  Gloves should be worn at all times during venipuncture to minimize exposure hazard. 

 
NOTE

:  VACUETTE

®

  QUICKSHIELD  Safety  Tube  Holders  are  designed  for  optimal  use with  VACUETTE

®

  Multiple  Use  Drawing  Needles 

with oriented needle bevels from Greiner Bio-One (except 18G). The use of needles from other manufacturers becomes the responsibility 
of the user.

 

 

Storage 

Store holders at

 

4

–36°C (40–97°F). 

NOTE

:  Avoid  exposure  to  direct  sunlight.  Exceeding  the  maximum  recommended  storage  temperature  may  lead  to  impairment  of  the 

holder. 

 
 

Handing Procedures 

 

READ THIS ENTIRE CIRCULAR BEFORE PERFORMING VENIPUNCTURE. 

WEAR  GLOVES  DURING  VENIPUNCTURE  AND  WHEN  HANDLING  BLOOD  COLLECTION  TUBES  TO  MINIMISE  EXPOSURE 
HAZARD! 
 
1.  Remove the cover from the valve section of the needle. 

NOTE

: The perforated label not only serves to simplify identification, but is also an indicator 

of  sterility  and  product  integrity.  If  the  perforations  are  broken  or  damaged,  dispose  of  the 
needle and select another. 
 

2.  Thread the needle perpendicularly into the VACUETTE

®

 QUICKSHIELD Safety Tube Holder. 

Ensure the needle is firmly seated so that it does not unthread during use. 

 
NOTE

:  Improper  threading  of  the  needle  into  the  VACUETTE

®

 

QUICKSHIELD  Safety  Tube 

Holder  can  result  in  damage  to  the  threads  of  the  holder  and/or  needle.  This  damage  can 
cause the needle to loosen during venipuncture. 

 
NOTE

:  Do not bend needles or use  VACUETTE

®

 

QUICKSHIELD Safety Tube Holder when 

needle is bent. 
 

3.  Select the puncture site. Apply the tourniquet (max. 1 minute) Prepare the  venipuncture site 

with  appropriate  antiseptic.  DO  NOT  PALPATE  THE  VENIPUNCTURE  AREA  AFTER 
CLEANSING! 

 

 

 

 

Отзывы: