Anaesthetic gas scavenging system AGSS
Greggersen Gasetechnik GmbH, Bodestraße 27-29, 21031 Hamburg, Germany, Fon +49 (0)40-73 93 57-0
04.12.13/GR/ AGFS-Forano.dts.eng.pdf
6
Technical data
Type:
All metal design, plug-in ejector for generating a vacuum, one-hand
operation for plugging in and out, visual signal for operating control,
integrated valve for capacity adjustment.
flush mounting/hollow wall: stainless steel installation housing, stainless
steel front plate with visual signal.
Connections:
Supply lines
pressure gas line:
Cu-tube 8x1
exhaust line:
Cu-tube 15x1
Outlet/ejector
pressure gas supply:
coupling for 4 mm PA-hose
or 6 mm hose nozzle
ejector:
coupling for 4 mm PA-hose
or 6 mm hose nozzle
anesthesia gas exhaust: 15 mm hose
visual signal:
coupling for 4 mm PA-hose
Capacity:
at least 50 l/min at propellant pressure of 500 kPa
Gasconsumption:
power gas: ≈12 l/min at flow 50 l/min
Color:
magenta-red designation, according to DIN EN ISO 9170-2
Installation:
•
flush-mounting or hollow wall mounting
•
ceiling supply systems
•
wall supply systems
dimensions see annex
Guidelines:
DIN EN ISO 9170-2
DIN EN ISO 7396-2
Medical Devices Directive 93/42/EEC; Classification: IIa
Important information
•
Comply with DIN EN ISO 7396-2 for the installation of AGSS systems!
•
Comply with TRGS 525 for operating the anesthesia gas exhaust system!
•
Use only original GREGGERSEN spare parts!
Service/Maintenance
External disinfection with a cloth is sufficient for all chrome-plated brass parts. Clean plastic parts
with a common disinfectant solution and follow manufacturers instructions. Let outlets and plug
dry completely before using them again.
Prior to each use, make a visual check of the outlets (external damages). Outlets and plug should
be controlled at least once a year (capacity, function); in very frequently used areas accordingly
more frequently!
Failures and damages may only be remedied by authorized expert personnel. For repairs, use only
original GREGGERSEN spare parts!