Commissioning 

 

Version 1.1 

– Revision Jan 2019 – Part nr.: 02156   

 

 

 

 

18

 

4

 

Commissioning 

4.1

 

Installation and Set-up Pre-checks 

After completing the installation and before starting the system for the first time, 
please check the following: 

 

Check and confirm  that  all electrical  connections  are carried out  according to 
the manuals and the technical drawings.

 

 

Check that the air supply is connected to the air supply filter regulator without 
leaking and in accordance with good installation practice.

 

 

Check  that  the  probe  is  installed  in  accordance  with  the  instructions  i.e.  in  a 
suitable  location  representing  the  gas  to  be  tested,  without  leaking  and  in  ac-
cordance with good installation practice.

 

 

Check the settings of the analyzer’s input and output signal. Please see the ana-
lyzer manual for the settings. 

 

Re-check all connections to make sure there is no air leaking. 

A leaking con-

nection will result in loss of air and may result in poor calibration

 

4.2

 

Start of Systems 

Using the G41 NO

X

/O

Analyzer manual, follow the instructions for commission-

ing of the analyzer and start the system. For calibration, follow the instructions in 
the section 5.1 of this manual and section 6 of the analyzer manual.  

 

  

Содержание G4130 NOX/O2

Страница 1: ...PART NUMBER 02156 G4130 NOX O2 ANALYZING SYSTEM WITH DIFFUSION PROBE MANUAL...

Страница 2: ......

Страница 3: ...Version 1 1 Revision Jan 2019 Part nr 02156 2...

Страница 4: ...FFUSION PROBE 12 3 5 ANALYZING BOARD 15 3 5 1 Mounting Panel 15 3 5 2 Air Supply Connection 15 3 5 3 Span NOX Gas Connection 15 3 5 4 Analyzer and Electrical Connections 15 4 COMMISSIONING 18 4 1 INST...

Страница 5: ...N LAYOUT OF THE DIFFUSION PROBE THREAD TYPE 11 FIGURE 3 3 DIFFUSION PROBE STANDARD WELDING TYPE 13 FIGURE 3 4 DIFFUSION PROBE SHORT WELDING TYPE 13 FIGURE 3 5 DIFFUSION PROBE STANDARD THREAD TYPE 14 F...

Страница 6: ...als must be read carefully in their entirety The manual has been made in general terms and do not take into consideration a specific installation As such the instruction manual is designed for the sta...

Страница 7: ...eedback All claims and inquiries for spares shall be addressed to Green Instruments A S or our distributors In all correspondence or when ordering spare parts please carefully state the equipment type...

Страница 8: ...proval Certificate Lloyd s Register Type Approval Certificate Diffusion Probe Sensor technology Heated zirconia type sensor Measurement range NOX 0 1500 ppm O2 0 0 21 Sample temperature 0 500 C Probe...

Страница 9: ...019 Part nr 02156 8 Umbilical cord Cord length 3 0 m Tubing In 28 mm nylon conduit Optional Equipment Remote display with alarm relays Recording and visualization Extension kit for umbilical cord Spec...

Страница 10: ...System are achieved by paying attention to the following points Make sure to install the probe in a suitable location where it will extract the gas that is going to be tested The probe shall be fitted...

Страница 11: ...hat it is easy accessible in connection with operation and service Exhaust Duct Distance to the next upstream bent min 3 duct diameters Distance to the engine turbo charger min 10 duct diamters Distan...

Страница 12: ...space 700 mm 600 mm 285 mm 600 285 180 mm Socket Yard supply to be welded Thread size 1 BSP Figure 3 2 Installation layout of the diffusion probe thread type 3 3 Safety Aspects Warning Make sure to fo...

Страница 13: ...which passes the probe must represent the gas to be tested and extracted from a location with laminar flow Holes and leaks in the gas ducting system before and after the probe may influence the accur...

Страница 14: ...G4130 NOx O2 Analyzing System 13 Figure 3 3 Diffusion probe standard welding type Figure 3 4 Diffusion probe short welding type...

Страница 15: ...System Installation Version 1 1 Revision Jan 2019 Part nr 02156 14 Figure 3 5 Diffusion probe standard Thread type Figure 3 6 Diffusion probe short Thread type...

Страница 16: ...ilter regulator has a filter and a drain The air supply connection to the filter regulator is a 1 8 BSP female connection The maximum pressure for the air supply inlet is 8 bar see Figure 3 8 3 5 3 Sp...

Страница 17: ...System Installation Version 1 1 Revision Jan 2019 Part nr 02156 16 Figure 3 7 Analyzing board dimensions...

Страница 18: ...G4130 NOx O2 Analyzing System 17 Figure 3 8 Piping and connection diagram...

Страница 19: ...t the probe is installed in accordance with the instructions i e in a suitable location representing the gas to be tested without leaking and in ac cordance with good installation practice Check the s...

Страница 20: ...r at the air supply filter regulator station The flow of zero NOX gas should be approximately 0 5 1 l min which can be adjust ed at the calibration flow meter For span calibration Turn the selector va...

Страница 21: ...point to the Operation posi tion The instrument air for back flushing is regulated by the solenoid valve which is controlled by the analyzer For back flushing the solenoid valve will be opened automat...

Страница 22: ...ally inspect and maintain the air flow system and connections may lead to imprecise analyzer readings and thus a mal function of the system Make sure that the instrument air that is supplied at the ai...

Страница 23: ...hrough the following steps Turn off the analyzer and disconnect the air supply Unscrew the nuts and remove the blue cover plate of the probe To disconnect the sensor electronic control unit ECU press...

Страница 24: ...e probe filter may be contaminated In that case manual cleaning of the probe is required Warning The analyzer must be turned off and the air supply must be disconnect ed before you start working with...

Страница 25: ...connect the air supply Unscrew the nuts and remove the blue cover plate of the probe Unbolt the 3 bolts at the sensor and filter flange Take the sensor and filter flange out and unscrew the threaded f...

Страница 26: ...the serial number of the analyzer which you can find on the label on the right side of the blue analyzer box Part No Consumable Parts Part Description The specific appearance of the spare parts is su...

Страница 27: ...pkg of 10 01248 G41 NOx O2 Analyzer 100 230 VAC 01250 Manual G41 NOX O2 Analyzer Software version 2 04 01819 Manual G41 NOX O2 Analyzer Software version 2 05 01254 Flow meter for calibration 1 l m 01...

Страница 28: ...nd air hoses 3 m 02200 Additional meters for 02199 Umbilical Cord per additional meter 01905 Probe filter 01913 Gasket for DP sensor filter flange OD82 02035 Flange gasket for DP thread Type OD100 020...

Страница 29: ...ic appearance of the spare parts is subject change without notice the function however will not change 33592 Signal amplifier 4 20 mA optional Other optional equipment e g flow alarm remote digital di...

Страница 30: ...29...

Страница 31: ......

Страница 32: ...Instruments USA Inc usa greeninstruments com 6750 N Andrews Avenue Suit 200 Fort Lauderdale FL 33309 USA Tel 1 954 613 0400 asia Green Instruments S Pte Ltd pua greeninstruments com 240 Woodlands Ind...

Отзывы: