Chapter 2
16
Micro-Ion E Module Instruction Manual 354011, Rev. A
If the analog module has a VCR type fitting or ConFlat flange, the module
will be properly grounded via the vacuum chamber connection.
If the analog or RS–232 module has a KF flange, refer to Figure 2-4 and
follow these instructions to ground the module to the vacuum chamber:
a.
Attach a metal hose clamp or other metal clamp to the gauge stem of
the housing.
b.
Install a 3.31 mm
2
(12 AWG) or larger copper wire between the clamp
and a metal ground lug, bolt, or stud on the vacuum chamber.
Figure 2-4
Vacuum Chamber Ground Connection
2.4
Eliminating Radio
Frequency Interference
The module has been tested and found to comply with U.S. Federal
Communications Commission (FCC) limits for a Class A digital device,
pursuant to Part 15 of the FCC rules. These limits provide reasonable
protection against harmful interference when the module operates in a
commercial environment.
The module can radiate radio frequency energy and, if not installed and
used in accordance with the instructions in this manual, may cause harmful
interference to other electrical equipment.
Metal hose clamp or
other metal clamp
3.31 mm
2
(12 AWG)
or larger ground wire
Vacuum Chamber
Ground lug, bolt, or stud
Gauge stem
Содержание Micro-Ion 354 E Series
Страница 2: ......
Страница 10: ...10 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 24: ...24 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 42: ...Chapter 6 42 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 46: ...46 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 52: ...52 Micro Ion E Module Instruction Manual 354011 Rev A...
Страница 53: ......